MR Confon Starter f MKII+ User Guide
071-CRS-N3001-R01
24
MRI Room Assembly
Follow the steps in this section to install the headphones.
Use only MR Safe components in the MRI Room.
Headphones
The red BNC connector corresponds to right audio channel, while the
white BNC connector corresponds to the left audio channel.
Figure 30 : BNC Connectors
Filter Elements
Filter elements are MR-conditional, meaning that they are safe when
installed properly in the MRI room. Make sure the filter elements are
securely fastened amongst themselves and onto the bulkhead BNC
connectors mounted on the RF Filter Panel.
Attach the headphone BNC connectors to the male coupler of the filter
elements. If the filter elements have been disassembled, they can be
reassembled as shown in Figure 31 or Figure 32: