Offset Zero Procedure

The CT4192 can be adjusted to zero the probe’s offset voltage 

using the offset adjustment tool supplied with the probe. Follow this 

procedure to perform the offset adjustment. 

1. 

Connect the probe to Channel 1 of the oscilloscope. Turn on the 

probe power. You may use the CT3723 or batteries to power the 

probe. Set the probe attenuation ratio to the low setting. 

2. 

Short the + and - probe inputs together with the hook tips.

3. 

Turn on power to the oscilloscope. Leave both the instrument 

and the probe turned on for 30 minutes to stabilize.

4. 

Press [Default Setup] and [Auto] on the oscilloscope.

5. 

Press the Channel 1 button, then press the probe softkey and 

set the attenuation to match that of the probe.

6. 

Set the oscilloscope to DC coupled mode and then the trace on 

the scope to 0 volts.

7. 

If possible, set the oscilloscope to average mode (x16) or high-

resolution mode to reduce oscilloscope noise.

8. 

Using the offset adjustment tool (included), adjust the probe 

offset voltage to 0 volts.

4  Cleaning

This probe does not require any particular cleaning. If necessary, clean 

the case with a soft cloth.

WARNING

Dry the probe thoroughly before attempting to make voltage 

measurements.

CAUTION

Avoid immersing or using abrasive cleaners or solvents containing 

Benzene (or similar solvents) on the probe as these can cause 

deterioration of the probe body and cables.  

Содержание CT4192

Страница 1: ...CT4192 User Manual...

Страница 2: ...ne Do not work alone when working with high voltages Inspect the Probe Inspect the probe and accessories for cracks and frayed or broken leads before each use If defects or damages are noted DO NOT US...

Страница 3: ...r Not for Critical Applications This probe is not authorized for use in contact with the human body or for use as a component in a life support device or system Do Not Substitute Parts Do not install...

Страница 4: ...ainst any unintended operation Compliance Statements Disposal of Old Electrical Electronic Equipment Applicable in the European Union and other European countries with separate collection systems This...

Страница 5: ...3dB z z Up to 700 V DC AC peak common mode z z Selectable attenuation settings of 10x 100x z z Power indicator z z High accuracy 1 z z Powered by 4 AA batteries included z z Power adapter CT3723 optio...

Страница 6: ...erview CT 4192 Description CT4192 1 M Hook probes black and red Attenuation Function Switch Output BNC Connector Power On Off Switch Power Indicator Red Offset Adjust External Power Connection Figure...

Страница 7: ...teries being sure to match polarity with the printed indicators inside the compartment Then Slide the cover back and replace the rubber boot See Figure 2 AA AA AA AA Figure 2 Battery Connection Altern...

Страница 8: ...V switch the attenuation ratio to 10x in order to get higher resolution and less noise ratio For voltages up to 700 V set the attenuation ratio to 100x 4 Plug the hook probes onto the test leads and c...

Страница 9: ...5 50 V div These values apply when the oscilloscope is set to the typical 1 M impedance input When the scope is set to 50 input the actual vertical scale will be doubled 10 V div for the 10x setting...

Страница 10: ...Default Setup and Auto on the oscilloscope 5 Press the Channel 1 button then press the probe softkey and set the attenuation to match that of the probe 6 Set the oscilloscope to DC coupled mode and t...

Страница 11: ...ntial Input Voltage DC AC peak 70 V 10x 700 V 100x Maximum Common Mode Input Voltage DC AC peak 70 V 10x 700 V 100x Absolute Max Rated Input Voltage each side to ground 1000 Vrms Input Resistance Capa...

Страница 12: ...de Operating 3 000 m Nonoperating 15 300 m Safety Specifications IEC 61010 031 CAT II Mechanical Characteristics Weight 400 g with probe and rubber boot Dimensions 170 x 63 x 21 mm BNC Cable Length 95...

Страница 13: ...Derating Curve The derating curve of the absolute maximum input voltage in common mode is shown as follows 1 000 10 000 100 10 1 10 100 Frequency MHz Volgage Vrms CT4192 Maximum Input Voltage Deratin...

Страница 14: ...persons will not be covered by this warranty Cal Test Electronics reserves the right to discontinue models change specifications price or design of this device at any time without notice and without...

Страница 15: ...450866 001 Rev B caltestelectronics com 14 Notes...

Страница 16: ...2017 Cal Test Electronics 450866 001 Rev B...

Отзывы: