16.4. Examples
The following section contains examples of commonly used operations involving the status
registers.
You can determine the reason for an MSS bit set by the following actions:
Step 1 :
Determine which summary bits are active. Use:
*STB?
Step 2 :
Read the corresponding Event register for each summary bit to determine which
events caused the summary bit to be set. Use:
ESR?
Note:
When an Event register is read, it is cleared. This also clears the
corresponding summary bit.
Step 3 :
Remove the specific condition that caused the event. If this is not possible, the
event may be disabled by programming the corresponding bit of the status group
Enable.
RP Series
October 2005
73