Warning:
Do not operate Hi Flow Stroller while
attached to the stationary unit.
Warning:
Should the leakage be excessive to the
point that a stream of liquid is present, leave the area
and call your health care provider immediately.
Caution:
If the Hi Flow Stroller is being refilled
immediately after a period of use, fill it, and then wait
10 minutes after filling with the FCV set at zero for
the pressure to stabilize. Then operate as normal.
Warning:
Liquid or cold gaseous oxygen can frostbite
the skin. Oxygen will result in increased fire hazard.
Portable Release
Button (push-on
style only)
Closing the valve at the first sign of liquid discharge
does not fill the Hi Flow Stroller as completely, but
saves liquid oxygen.
9. For the push-on style connectors, push the release
button down until the units separate. For all other
connectors rotate the unit until it separates from the
stationary unit.
Caution:
Check liquid level gauge only after the vent
valve is closed.
Caution:
If the Hi Flow Stroller does not separate
easily, do not use force. The units may be frozen
together. Wait until the units warm up — then they’ll
separate easily. Do not touch any frosted parts.
Should there be any liquid leakage from the
stationary unit or Hi Flow Stroller after separation,
reattach the Hi Flow Stroller immediately. Again
separate the units. If the problem persists, leave
units coupled and call your health care provider.