10 - ENG    PN 21033264 B | User Manual

HELiOS 300 & 800

User Troubleshooting

The following information is intended to help you 

troubleshoot and solve simple operational problems 

that you may experience when using your HELiOS 

Portable.

1. The Portable makes a hissing sound.

• Hissing can occur to maintain the correct 

operating pressure within the Portable. It is 

most likely to hiss after filling or when the 

position of the Portable is changed. Hissing 

can last for approximately 10 minutes after 

filling. Additionally, it can occur when the 

flow control valve is at a low setting.

2. The Portable does not pulse after filling.

• If during the filling process, the Portable is 

disengaged with the vent valve lever down, 

pressure may be reduced within the Portable 

causing a delay in the conserving device 

function. It may require as much as 60 min-

utes to restore adequate pressure for accurate 

oxygen flow. Additionally, improper filling 

or lower than normal operating pressure in 

the Reservoir will contribute to the unit not 

pulsing.

3. The Portable stops pulsing during use.

• Ensure that both connections of the dual 

lumen cannula are firmly attached to the 

Oxygen Outlet and Sensor connectors.

• Change the cannula if water droplets are pres-

ent from humidified exhaled gas. Replacing 

the cannula with a dry one enhances the 

ability of the Portable to pulse.

• Ensure that the cannula is not kinked.

• Ensure that the cannula tips remain in your 

nostrils and do not slide to one side.

• Ensure that there is oxygen in the Portable.

• With the cannula on, close your mouth and 

breathe only through your nose to verify that 

the Portable has stopped pulsing.

4. The Portable does not fill.

• Verify that there is oxygen in the Reservoir.

• Ensure that the Portable and Reservoir fill 

connectors are fully engaged throughout the 

filling process.

5. The Portable vent valve does not close properly 

at the end of the filling process.

• If the vent valve fails to close and the hissing 

sound and oxygen vapor cloud continue, 

carefully remove the Portable by depressing 

the release button on the  Reservoir. Venting 

from the bottom of the Portable will stop in 

a few minutes. Allow the unit to warm until 

you can close the vent valve. The Portable 

may require as much as 60 minutes to restore 

Operating Instructions –  
Level Verification

1. Lift the H300/H850 by the contents indicator 

strap.

2. Push the bottom backside of the unit forward 

until it is straight up and down.

3. Observe the green bar in the contents indica-

tor window.  The window is completely filled 

with the green bar when the unit is full of liquid 

oxygen.

Portable Full 

              Portable Empty 

  

Checking the H300/H850 contents indicator

Содержание HELiOS Marathon

Страница 1: ...HELiOSTM Series User Manual English...

Страница 2: ...prohibited Reg P003 Do not smoke near unit or while operating unit Reg P002 Type BF applied part degree of protection against electric shock Reg 5333 Warning Reg W001 Council Directive 93 42 EEC conc...

Страница 3: ...ctive 2012 19 EU waste electrical and electronic equipment WEEE WEEE IEC 60601 1 Medical electrical equipment Part 1 General requirements for basic safety and essential performance Drip Proof This pro...

Страница 4: ...AUTIOUSLY RETURN THE UNIT INTO AN UPRIGHT POSITION IF POSSIBLE IF ANY LIQUID OXYGEN IS ESCAPING LEAVE THE AREA IMME DIATELY AND CALL YOUR HEALTH CARE PROVIDER DO NOT ATTEMPT TO MOVE THE UNIT OR STOP T...

Страница 5: ...d as a conserving or continuous flow device and holds 0 84 liters of liquid oxygen It provides continuous flow from 1 6 LPM and demand flow on settings of 1 4 H300 and H850 portable units shown Specif...

Страница 6: ...rap 5 Contents Indicator Window 6 Vent Valve Lever 7 Fill Connector QDV Fill Connector Compatibility The H300 and H850 can be filled from any CAIRE top fill TF reservoir that has a CAIRE style male fi...

Страница 7: ...through top port Breath is detected through bottom port Operating Instructions Filling 1 Clean the fill connectors on the H300 H850 and the reservoir with a clean dry lint free cloth WARNING THE FILL...

Страница 8: ...T LEAVE THE AREA AND CALL YOUR HEALTH CARE PROVIDER IMMEDIATELY WARNING IF LARGE AMOUNTS OF VAPOR ARE ISSUING FROM THE UNITS DURING FILLING STOP FILLING LEAVE THE ROOM AND CALL YOUR HEALTH CARE PROVID...

Страница 9: ...continuous flow the H300 H850 will deliver a steady stream of oxygen You must continue to breathe through the nasal cannula normally to continue receiving continuous flow Note Due to the possibility...

Страница 10: ...pres ent from humidified exhaled gas Replacing the cannula with a dry one enhances the ability of the Portable to pulse Ensure that the cannula is not kinked Ensure that the cannula tips remain in you...

Страница 11: ...r always wipe the male fill connector on the Reservoir and the female fill connector on the Portable with a clean dry cloth before filling NOTE It is recommended that routine maintenance be performed...

Страница 12: ...with no further obligation or consequence All rights not expressly stated herein are reserved by us as applicable www cairemedical com CAIRE Inc 2200 Airport Industrial Dr Ste 500 Ball Ground GA 3010...

Отзывы: