12 - ENG PN MN232-C4 J | User Manual
FreeStyle Comfort
5. Turn on the FreeStyle Comfort by pressing the
power button and select your prescribed flow
setting utilizing the increase (+) or decrease (-)
flow buttons.
A brief audible alarm will sound to indicate the
device has powered on.
Note: When the FreeStyle Comfort is initially turned
on the default flow setting is 2. If the setting is
changed, the unit will power on to the previous
flow setting.
6. Position the nasal cannula on your face per the
manufacturer’s instructions provided with the
cannula and breathe normally. You should feel
a puff of oxygen delivered through the nasal
cannula each time you inhale.
7. To change the flow setting, press the increase (+)
or decrease (-) flow buttons to the desired flow
rate. It is normal to hear a difference in sound as
you change the flow settings.
autoSAT Pulse Volume (mL)*Setting
Breath
Rate
1
2
3
4
5
15
10.5
21.0
31.5
42.0
52.5
20
10.5
21.0
31.5
42.0
52.5
25
10.5
21.0
31.5
42.0
42.0
30
10.5
21.0
31.5
35.0
35.0
35
10.5
21.0
30.0
30.0
30.0
40
10.5
21.0
26.3
26.3
26.3
*Tolerance is +/- 15%
non-autoSAT Pulse Volume (mL)* Setting
Breath
Rate
1
2
3
4
5
15
14.0
28.0
42.0
56.0
70.0
20
10.5
21.0
31.5
42.0
52.5
25
8.4
16.8
25.2
33.6
42.0
30
7.0
14.0
21.0
28.0
35.0
35
6.0
12.0
18.0
24.0
30.0
40
5.3
10.5
15.8
21.0
26.3
8. Turn the FreeStyle Comfort off by pressing the
power button.
Note: The display screen will enter standby mode
and dim after 20 seconds. Press any button to wake
the screen out of standby mode.
9. For Advanced Features see Menu Navigation
Instructions
SPECIAL PRODUCT FEATURES
autoSAT
®
CAIRE’s proprietary autoSAT feature allows the
FreeStyle Comfort to deliver a consistent volume
of oxygen (bolus) per breath, up to a maximum
breath rate. As the user’s breath rate increases
during activities like walking, the autoSAT feature
adjusts the device operation to maintain a set bolus
volume.
UltraSense
®
This sensitive pulse dose trigger detects the
pressure change during your breath, and releases
the pulse of oxygen when you inhale.
autoDOSE™
Should a “no breath alarm” occur, this feature will
pulse oxygen at 20 breaths per minute at the unit’s
current flow setting until the alarm state is resolved.
Содержание FreeStyle Comfort
Страница 345: ...FreeStyle Comfort PN MN232 C4 J GRE 345 FreeStyle Comfort FreeStyle Comfort FreeStyle Comfort 8 AC AC DC 8 16 carry all...
Страница 346: ...FreeStyle Comfort 346 GRE PN MN232 C4 J AC DC autoSAT...
Страница 347: ...FreeStyle Comfort PN MN232 C4 J GRE 347 FreeStyleComfort FREESTYLE COMFORT 1 2...
Страница 348: ...FreeStyle Comfort 348 GRE PN MN232 C4 J DOT UN AC GFI...
Страница 349: ...FreeStyle Comfort PN MN232 C4 J GRE 349 3 10 WWW P65WARNINGS CA GOV AC AC...
Страница 350: ...FreeStyle Comfort 350 GRE PN MN232 C4 J DC DC RF 25 50 20 C 20 C 4 F 68 F AC DC...
Страница 358: ...FreeStyle Comfort 358 GRE PN MN232 C4 J FreeStyle Comfort FreeStyle Comfort 1 FreeStyle Comfort 2 3 4 70 50 75 LED...
Страница 359: ...FreeStyle Comfort PN MN232 C4 J GRE 359 AC DC 3 6 3...
Страница 360: ...FreeStyle Comfort 360 GRE PN MN232 C4 J 1 1 20 2 2 1 20 2 3 1 20 2 4 1 30 15 5 1 30 6...
Страница 361: ...FreeStyle Comfort PN MN232 C4 J GRE 361 7 1 30 5 8 OCD 1 30 9 1 10 1 60 40bpm 11 12 13 1...
Страница 362: ...FreeStyle Comfort 362 GRE PN MN232 C4 J autoSAT DC autoSAT DC 1 CAIRE 2 5 1 CAIRE 2 1...
Страница 363: ...FreeStyle Comfort PN MN232 C4 J GRE 363 5 1 1...
Страница 367: ...FreeStyle Comfort PN MN232 C4 J GRE 367 FreeStyle Comfort AC DC AC DC AC DC...
Страница 498: ...498 ENG PN MN232 C4 J User Manual FreeStyle Comfort Notes...
Страница 499: ...PN MN232 C4 J User Manual ENG 499 FreeStyle Comfort Notes...