Recommended separation distances between portable and mobile RF
communications equipment and the FreeStyle or FreeStyle 5
The FreeStyle and FreeStyle 5 are intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the FreeStyle and FreeStyle 5 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the FreeStyle and FreeStyle 5 as recommended below, according
to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
P
d
2
.
1
=
80 MHz to 800 MHz
P
d
2
.
1
=
800 MHz to 2,5 GHz
P
d
2
.
1
=
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Classification
Type of protection against electric shock:
Class II
Protection from electric shock is
achieved by double insulation.
Degree of protection against electric shock:
Type BF Equipment providing a particular
degree of protection against electric shock
regarding
1) allowable leakage current;
2) reliability of protective earth connection
(if present).
Not intended for direct cardiac application.
Independent testing for Medical Electrical
Equipment Standard:
Tested to be in compliance with
• IEC 60601-1 General Requirements for
safety.
• CAN/CSA C22.2 No. 60601-1-08 M90
Medical Electrical Equipment – Part 1:
General Requirements for Safety
Protection against potential electromagnetic or
other interference between the equipment and other
devices.
• Tested to be in compliance with EN60601-1-2
(EMC).
• Tested to be in compliance with RTCA/DO160
Section 21 Category M.
• CISPR 11 / EN 55011 Class B Group 1,
“Industrial, Scientific, and Medical (ISM)
Equipment”
• FCC Part 15, Subpart B – Class B
Unintentional Radiators
Method of cleaning and infection control allowed:
Please refer to “Cleaning, Care, and Proper
Maintenance” section of this FreeStyle User
Manual.
Degree of safety of application in the presence of
flammable anesthetic gases:
Equipment not suited for such application.
Mode of operation:
Continuous duty.
Содержание FreeStyle 5
Страница 323: ...Freestyle FreeStyle 5 323 21 78 1 FreeStyle FreeStyle 5 FreeStyle FreeStyle 5 21 FreeStyle FreeStyle 5 AC DC...
Страница 324: ...324 MN218 C4 A Greek POC AC DC 65 FreeStyle FreeStyle 5 FreeStyle carry all AC DC AC DC FreeStyle 5 carry all AC DC...
Страница 325: ...Freestyle FreeStyle 5 325 FreeStyle FreeStyle 5 FreeStyle FreeStyle 5...
Страница 326: ...326 MN218 C4 A Greek 25...
Страница 327: ...Freestyle FreeStyle 5 327 FreeStyle FreeStyle5 1 2 3...
Страница 328: ...328 MN218 C4 A Greek DOT UN AC GFI...
Страница 329: ...Freestyle FreeStyle 5 329 3 FREESTYLE FREESTYLE5...
Страница 330: ...330 MN218 C4 A Greek AC AC DC DC DC DC DC DC DC RF 25 50 23 C 2 C 73 F 35 F FREESTYLE FREESTYLE5 AC DC...
Страница 332: ...332 MN218 C4 A Greek 1 FreeStyle FreeStyle 5 a b c 2 a FreeStyle FreeStyle 5 b 3 AC DC 4 7 62m 25ft...
Страница 333: ...Freestyle FreeStyle 5 333 5 FreeStyle FreeStyle 5 6 7 8 FreeStyle FreeStyle 5...
Страница 343: ...Freestyle FreeStyle 5 343 FreeStyle DC AC FreeStyle 15 BATT DC AC 2 BATT FreeStyle FreeStyle DC AC FreeStyle DC DC DC DC...
Страница 345: ...Freestyle FreeStyle 5 345 FreeStyle FreeStyle 5 1 2 1 2 3...