30 — MN218 B
Recommended separation distances between portable and mobile RF
communications equipment and the FreeStyle or FreeStyle 5
The FreeStyle and FreeStyle 5 are intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the FreeStyle and FreeStyle 5 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the FreeStyle and FreeStyle 5 as recommended below, according
to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
P
d
2
.
1
80 MHz to 800 MHz
P
d
2
.
1
800 MHz to 2,5 GHz
P
d
2
.
1
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Classi
fi
cation
Type of protection against electric shock:
Class II
Protection from electric shock is
achieved by double insulation.
Degree of protection against electric shock:
7\SH%) (TXLSPHQWSURYLGLQJDSDUWLFXODU
degree of protection against electric
shock regarding
1) allowable leakage current;
2) reliability of protective earth connection
(if present).
Not intended for direct cardiac application.
Independent testing for Medical Electrical
(TXLSPHQW6WDQGDUG
Tested to be in compliance with
,(&*HQHUDO5HTXLUHPHQWVIRU
safety.
&$1&6$&1R0
0HGLFDO(OHFWULFDO(TXLSPHQW±3DUW
*HQHUDO5HTXLUHPHQWVIRU6DIHW\
Protection against potential electromagnetic or
RWKHULQWHUIHUHQFHEHWZHHQWKHHTXLSPHQWDQGRWKHU
devices.
7HVWHGWREHLQFRPSOLDQFHZLWK(1
(EMC).
• Tested to be in compliance with RTCA/DO160
Section 21 Category M.
• CISPR 11 / EN 55011 Class B Group 1,
³,QGXVWULDO6FLHQWL¿FDQG0HGLFDO,60
(TXLSPHQW´
• FCC Part 15, Subpart B – Class B
Unintentional Radiators
Method of cleaning and infection control allowed:
Please refer to “Cleaning, Care, and Proper
Maintenance” section of this FreeStyle User
Manual.
Degree of safety of application in the presence of
ÀDPPDEOHDQHVWKHWLFJDVHV
(TXLSPHQWQRWVXLWHGIRUVXFKDSSOLFDWLRQ
Mode of operation:
Continuous duty.