24 — MN218 B
Troubleshooting
Problem
Probable Cause
Solution
FreeStyle does not operate when
a pulse flow selection button is
pressed.
Battery is discharged.
Malfunction.
Power the unit through the DC outlet, or an AC
outlet.
Contact your Equipment Provider, and change to
another source of oxygen as necessary.
A continuous alarm sounds
and the (yellow) alarm light
illuminates continuously.
FreeStyle has not detected a
breath for 15 minutes.
Check the cannula connection.
Ensure that cannula tubing is not kinked.
Make sure that the cannula is positioned properly
and that you are breathing through your nose. For
mouth-breathing your clinician may recommend a
chin strap.
Intermittent alarm condition,
and the (yellow) BATT light
illuminates intermittently.
Battery requires charging.
Connect to a DC or an AC outlet within 2 minutes, or
exchange battery packs for a fully charged pack.
Rapid alarm condition, and the
(yellow) BATT light illuminates
intermittently and FreeStyle
shuts down.
Battery pack charge is too low to
operate the FreeStyle unit.
Connect to DC or an AC outlet immediately.
Intermittent alarm condition,
and the (red) alarm light
illuminates intermittently.
Breathing rate has exceeded the
capacity of the FreeStyle unit
Reduce activity, and then turn unit off and back on
again to reset unit. If necessary, change to another
source of oxygen as available and contact your
Equipment Provider.
Intermittent alarm condition,
and the (red) alarm light
illuminates continuously.
A general malfunction has
occurred.
Change to another source of oxygen as available, and
contact your Equipment Provider.
Unit does not start on battery
pack power.
Unit may be hot or cold if
left outdoors such as in an
automobile.
Allow the unit to reach normal operating
temperature, which may take several minutes if
exposed to temperature extremes.
Delay in recharging internal
battery.
Internal battery exceeds charging
temperature.
Unit may be operated; however, charging may not
resume until battery temperature is reduced.
Unit alarms while in automobile
and connected to the DC outlet.
No power to the unit if battery
depleted and DC outlet not
charging.
Disconnect the power supply from the automobile
outlet, restart the automobile, and then reconnect
the power supply into the automobile DC outlet to
reset the breaker within DC power supply.
All other problems.
Change to another source of oxygen as available, and
contact your Equipment Provider.