UM2679 – A1536 AG536 Family Power Supply Boards 

11 

OVV 

(channel in OVERVOLTAGE condition) 

UNV 

(channel in UNDERVOLTAGE condition)5 

EXTTRIP 

(channel OFF due to external TRIP line signal)6 

INTTRIP 

(channel OFF due to internal OVERCURRENT condition) 

EXT_DIS 

(channel disabled by board INTERLOCK protection) 

Moreover it is possible to monitor board temperature and to check board status; the following messages may be 
returned by the SYSTEM when monitoring the board status: 

UNDER_TEMP    (board temperature < 5°C ) 
OVER_TEMP       (board temperature > 65°C) 

Output Enable  

In order to enable the HV output channels, first of all it is necessary that pin 9 and 10 on the Radial 52pin output 
connector are short circuited (see p.7); if the board features SHV connectors, skip this step. 
Then the enable procedure is completed in one of the following ways: 

 

Boards with LEMO interlock connectors 

 

 

 

terminating the PASSIVE INTERLOCK [P] (see External connections) connector on 50 Ohm. 

 

supplying the SIGNAL INTERLOCK [S] (see External connections) connector with a +5 V (3-4mA) signal. 

 

Boards with AMP interlock connectors 

 

 

Short circuit pin 3 and 4 [the lower ones]. Leave contact open between pin 1 (+) and 2 (-) [upper ones] 

 

Apply +5 V (3-4mA) differential signal between pin 1 (+) and 2 (-); pin 1 is the upper one. Leave contact open 
between pin 3 and 4. 

 
The INTERLOCK LED (red) is turned off as one of the actions above is performed. 
When the channels are disabled the voltage outputs drop to zero at the maximum rate available; when the output 
disable cause is removed (see above), the channels remain OFF until the User turns them ON via software. 

Grounding specifications 

The A1536 channels  share a common floating return (RTN), insulated from the crate ground (AGND/EARTH). This 
feature allows on-detector grounding, thus avoiding loops which may increase noise level. RTN and AGND may be 
coupled in several ways, according to environment requirements. Examples refer to 32 channel version (A1536). 

Safety Earth connection 

The connection of shield and return to Earth is fundamental for User safety. The connection must always be at the 
level  of  detector  or  power  supply  system.  Return  and  Shield  connections,  even  if  not  present  or  performed 
incorrectly,  due  to  protection  circuits  implemented  on  the  A1536  are  bound  to  Earth;  in  this  case  the  voltage 
difference between return and Earth (System), shield and Earth is limited to approximately 50V.  Please note that 
this  is  a  status  of  emergency-protection,  not  a  working  one.  The  Connector  Configurator  allows  to  optimize  the 

                                                                        
 
 

5

 

UNV is also reported when Hvmax limit is reached, it is up to the User to verify that VMON value does not exceed HVMAX.

 

6

 EXTTRIP and INTTRIP parameters are expressed in Hexadecimal format 

Содержание A1536

Страница 1: ...CAEN Tools for Discovery n Electronic Instrumentation Rev 10 Es A1536 AG536 Family Power Supply Boards Rev 10 11 February 2015 ...

Страница 2: ...replace any product within the guarantee period if the Guarantor declares that the product is defective due to workmanship or materials and has not been caused by mishandling negligence on behalf of the User accident or any abnormal conditions or operations CAEN declines all responsibility for damages or injuries caused by an improper use of the Modules due to negligence on behalf of the User It i...

Страница 3: ...tion requirements 9 General safety information 9 Injury Precautions 9 Safety Terms and Symbols on the Product 9 Installation 9 3 Operating modes 10 Output control and monitoring 10 Output Enable 11 Grounding specifications 11 Safety Earth connection 11 List of Figures Fig 1 A1536 A1536D and A1536S front panel 6 Fig 2 JA jumper location 8 Fig 3 Earth configuration connection examples 12 List of Tab...

Страница 4: ...ons are single width 5 TE Multipin versions have also the safety board interlock this protection allows to disable the primary HV generation when the HV outputs are not connected to their loads Tab 1 Available versions Version Ch Nr Polarity Connector Return Size IMax A1536DN 12 Negative SHV Common Floating 1U 1 mA A1536DP 12 Positive SHV Common Floating 1U 1 mA A1536DM 12 Mixed SHV Common Floatin...

Страница 5: ...ec 1 Volt sec step Voltage Ripple 1 15 mV typical 30mV max 10 mV typical 15mV max VMon vs VOut Accuracy 2 3 typical 0 3 0 25 V VSet vs VOut Accuracy 2 typical 0 3 2 5 V IMon vs IOut Accuracy 2 typical 2 500 nA High Power typical 2 50 nA High Resolution typical 2 5 nA ISet vs IOut Accuracy 2 typical 2 1µA Maximum output power 3 W per channel 1 5W per channel Power consumption 160 W 120 W 60W full p...

Страница 6: ...CAEN Electronic Instrumentation UM2679 A1536 AG536 Family Power Supply Boards 6 Front Panel Fig 1 A1536 A1536D and A1536S front panel ...

Страница 7: ...nt Table 2 52 pin connector assignment 1 N C 11 Return 22 N C 32 Return 43 N C 2 N C 12 N C 23 N C 33 N C 44 N C 3 HVOUT23 13 N C 24 HVOUT12 34 HVOUT6 45 HVOUT0 4 HVOUT24 14 HVOUT18 25 HVOUT13 35 HVOUT7 46 HVOUT1 5 HVOUT25 15 HVOUT19 26 HVOUT14 36 HVOUT8 47 HVOUT2 6 HVOUT26 16 HVOUT20 27 HVOUT15 37 HVOUT9 48 HVOUT3 7 HVOUT27 17 HVOUT21 28 HVOUT16 38 HVOUT10 49 HVOUT4 8 HVOUT28 18 HVOUT22 29 HVOUT1...

Страница 8: ... 1 2 Agnd Shield Connects Agnd Earth to HV cable shield A1536 A1536L 3 4 Agnd Return Connects Agnd Earth to HV channels return A1536 A1536L A1536S 5 6 Shield Return Connects Shield to HV channels return A1536 A1536L On the 1536D RTN and AGND Earth may be connected by short circuiting JA jumper pins on the motherboard see figure below The protection shield must be screwed off in order to access JA ...

Страница 9: ...ut Covers To avoid electric shock or fire hazard do not operate this product with covers or panels removed Do Not Operate in Wet Damp Conditions To avoid electric shock do not operate this product in wet or damp conditions Do Not Operate in an Explosive Atmosphere To avoid injury or fire hazard do not operate this product in an explosive atmosphere Do Not Operate With Suspected Failures If you sus...

Страница 10: ...e OFF POn EN DIS Power On option which can be enabled or disabled If this option is enabled at Power On or after a Restart each channel is restored in the same condition defined by the Power parameter it was before the Power Off or Reset If this option is disabled at Power On or after a Restart all the channels are off independently from the condition in which they were before the Power Off or Res...

Страница 11: ...n 1 is the upper one Leave contact open between pin 3 and 4 The INTERLOCK LED red is turned off as one of the actions above is performed When the channels are disabled the voltage outputs drop to zero at the maximum rate available when the output disable cause is removed see above the channels remain OFF until the User turns them ON via software Grounding specifications The A1536 channels share a ...

Страница 12: ...ATION HV Generator Load PMT 0 HV Channel 0 RETURN Load PMT 31 HV Channel 31 EARTH of DETECTOR Cable SHIELD Circuit Protected SHIELD Circuit Protected RETURN HV CABLE AGND 2 1 4 3 6 5 FRONT PANEL JUMPERS CLOSED LOOP EARTH CONFIGURATION WITH PROTECTION STAGE ON LOAD GND HV Generator Load PMT 0 HV Channel 0 RETURN Load PMT 31 HV Channel 31 EARTH of DETECTOR Cable SHIELD Circuit Protected SHIELD Circu...

Страница 13: ...cluding engineers scientists and technical professionals who all trust them to help achieve their goals faster and more effectively CAEN S p A CAEN GmbH CAEN Technologies Inc Via Vetraia 11 Klingenstraße 108 1140 Bay Street Suite 2 C 55049 Viareggio D 42651 Solingen Germany Staten Island NY 10305 Italy Phone 49 0 212 254 4077 USA Tel 39 0584 388 398 Fax 49 0 212 25 44079 Tel 1 718 981 0401 Fax 39 ...

Отзывы: