background image

CAEN

 

 

Electronic Instrumentation

 

 

UM2679 – A1536 AG536 Family Power Supply Boards 

Technical Specifications 

Packaging 

24 channel SHV versions (

A1536S,  AG536S

) are double width boards (10 TE); all other versions are single width (5 TE); 

height is 6U. 

External connections 

The  function  and  electro-mechanical  specifications  of  the  external  connectors  are  listed  in  the  following 
subsections. 

Version: 

A1536, A1536L, AG536, AG536L 

A1536S, A1536D, AG536S, AG536D 

Output Channels: 

Multipin connector  Radiall 691803004 type, 52 pin 
male  (to  be  mated  with  Radiall  691802002  [

SCEM 

09.41.34.700.2

]

 

type

4

); see Table below  

HV 

coaxial 

connectors 

Radiall 

SHVR317580 

Return: 

Radiall R921921 socket, Ø 2mm (A1536x only) 

Not present on A1536D AG536D 

AGND: 

Radiall R921921 socket, Ø 2mm (A1536x only) 

Not present on A1536D AG536D 

PASSIVE INTERLOCK: 

00-type LEMO connector 

AMP 280371-2 (A1536D, AG536D) 

SIGNAL INTERLOCK: 

00-type LEMO connector 

AMP 280371-2 (A1536D, AG536D) 

Multipin connector pin assignment 

Table 2 – 52 pin connector assignment 

N.C. 

11  Return 

22  N.C. 

32  Return 

43 

N.C. 

N.C. 

12  N.C. 

23  N.C. 

33  N.C. 

44 

N.C. 

HVOUT23  13  N.C. 

24  HVOUT12 

34  HVOUT6 

45 

HVOUT0 

HVOUT24  14  HVOUT18 

25  HVOUT13 

35  HVOUT7 

46 

HVOUT1 

HVOUT25  15  HVOUT19 

26  HVOUT14 

36  HVOUT8 

47 

HVOUT2 

HVOUT26  16  HVOUT20 

27  HVOUT15 

37  HVOUT9 

48 

HVOUT3 

HVOUT27  17  HVOUT21 

28  HVOUT16 

38  HVOUT10 

49 

HVOUT4 

HVOUT28  18  HVOUT22 

29  HVOUT17 

39  HVOUT11 

50 

HVOUT5 

INT_A 

19  HVOUT29 

30  HVOUT31 

40  N.C. 

51 

N.C. 

10  INT_B 

20  HVOUT30 

31  SHIELD 

41  N.C. 

52 

SHIELD 

 

 

21  Return 

 

 

42  Return 

 

 

HV OUT 24 through 31 are “N.C.” on A1536L and AG536L. 

Other components 

HV ON LED: 

Function:

 lights up as at least one channel is on

  

Type: 

red  LEDs  for  positive  polarity  version

yellow  green  LEDs  for  negative  polarity 

version.                                             

INTERLOCK LED: 

Function:

 lights up as the board is in INTERLOCK (channel are disabled).    

 

Type: 

red LED                                       

 

VMAX trimmer: 

Function: 

it allows to adjust the hardware maximum voltage VMAX common to all the 

channels. Its value can be read out via software.

 

 

 

 

                                                                        
 
 

4

 

Requires 52 pins Radiall 691804300 [

SCEM 09.41.33.830.7

]

 

type, to be inserted using the insertion/extraction tool 

Radiall 282549024 [

SCEM 34.95.17.125.3

] type.

 

Содержание A1536

Страница 1: ...CAEN Tools for Discovery n Electronic Instrumentation Rev 10 Es A1536 AG536 Family Power Supply Boards Rev 10 11 February 2015 ...

Страница 2: ...replace any product within the guarantee period if the Guarantor declares that the product is defective due to workmanship or materials and has not been caused by mishandling negligence on behalf of the User accident or any abnormal conditions or operations CAEN declines all responsibility for damages or injuries caused by an improper use of the Modules due to negligence on behalf of the User It i...

Страница 3: ...tion requirements 9 General safety information 9 Injury Precautions 9 Safety Terms and Symbols on the Product 9 Installation 9 3 Operating modes 10 Output control and monitoring 10 Output Enable 11 Grounding specifications 11 Safety Earth connection 11 List of Figures Fig 1 A1536 A1536D and A1536S front panel 6 Fig 2 JA jumper location 8 Fig 3 Earth configuration connection examples 12 List of Tab...

Страница 4: ...ons are single width 5 TE Multipin versions have also the safety board interlock this protection allows to disable the primary HV generation when the HV outputs are not connected to their loads Tab 1 Available versions Version Ch Nr Polarity Connector Return Size IMax A1536DN 12 Negative SHV Common Floating 1U 1 mA A1536DP 12 Positive SHV Common Floating 1U 1 mA A1536DM 12 Mixed SHV Common Floatin...

Страница 5: ...ec 1 Volt sec step Voltage Ripple 1 15 mV typical 30mV max 10 mV typical 15mV max VMon vs VOut Accuracy 2 3 typical 0 3 0 25 V VSet vs VOut Accuracy 2 typical 0 3 2 5 V IMon vs IOut Accuracy 2 typical 2 500 nA High Power typical 2 50 nA High Resolution typical 2 5 nA ISet vs IOut Accuracy 2 typical 2 1µA Maximum output power 3 W per channel 1 5W per channel Power consumption 160 W 120 W 60W full p...

Страница 6: ...CAEN Electronic Instrumentation UM2679 A1536 AG536 Family Power Supply Boards 6 Front Panel Fig 1 A1536 A1536D and A1536S front panel ...

Страница 7: ...nt Table 2 52 pin connector assignment 1 N C 11 Return 22 N C 32 Return 43 N C 2 N C 12 N C 23 N C 33 N C 44 N C 3 HVOUT23 13 N C 24 HVOUT12 34 HVOUT6 45 HVOUT0 4 HVOUT24 14 HVOUT18 25 HVOUT13 35 HVOUT7 46 HVOUT1 5 HVOUT25 15 HVOUT19 26 HVOUT14 36 HVOUT8 47 HVOUT2 6 HVOUT26 16 HVOUT20 27 HVOUT15 37 HVOUT9 48 HVOUT3 7 HVOUT27 17 HVOUT21 28 HVOUT16 38 HVOUT10 49 HVOUT4 8 HVOUT28 18 HVOUT22 29 HVOUT1...

Страница 8: ... 1 2 Agnd Shield Connects Agnd Earth to HV cable shield A1536 A1536L 3 4 Agnd Return Connects Agnd Earth to HV channels return A1536 A1536L A1536S 5 6 Shield Return Connects Shield to HV channels return A1536 A1536L On the 1536D RTN and AGND Earth may be connected by short circuiting JA jumper pins on the motherboard see figure below The protection shield must be screwed off in order to access JA ...

Страница 9: ...ut Covers To avoid electric shock or fire hazard do not operate this product with covers or panels removed Do Not Operate in Wet Damp Conditions To avoid electric shock do not operate this product in wet or damp conditions Do Not Operate in an Explosive Atmosphere To avoid injury or fire hazard do not operate this product in an explosive atmosphere Do Not Operate With Suspected Failures If you sus...

Страница 10: ...e OFF POn EN DIS Power On option which can be enabled or disabled If this option is enabled at Power On or after a Restart each channel is restored in the same condition defined by the Power parameter it was before the Power Off or Reset If this option is disabled at Power On or after a Restart all the channels are off independently from the condition in which they were before the Power Off or Res...

Страница 11: ...n 1 is the upper one Leave contact open between pin 3 and 4 The INTERLOCK LED red is turned off as one of the actions above is performed When the channels are disabled the voltage outputs drop to zero at the maximum rate available when the output disable cause is removed see above the channels remain OFF until the User turns them ON via software Grounding specifications The A1536 channels share a ...

Страница 12: ...ATION HV Generator Load PMT 0 HV Channel 0 RETURN Load PMT 31 HV Channel 31 EARTH of DETECTOR Cable SHIELD Circuit Protected SHIELD Circuit Protected RETURN HV CABLE AGND 2 1 4 3 6 5 FRONT PANEL JUMPERS CLOSED LOOP EARTH CONFIGURATION WITH PROTECTION STAGE ON LOAD GND HV Generator Load PMT 0 HV Channel 0 RETURN Load PMT 31 HV Channel 31 EARTH of DETECTOR Cable SHIELD Circuit Protected SHIELD Circu...

Страница 13: ...cluding engineers scientists and technical professionals who all trust them to help achieve their goals faster and more effectively CAEN S p A CAEN GmbH CAEN Technologies Inc Via Vetraia 11 Klingenstraße 108 1140 Bay Street Suite 2 C 55049 Viareggio D 42651 Solingen Germany Staten Island NY 10305 Italy Phone 49 0 212 254 4077 USA Tel 39 0584 388 398 Fax 49 0 212 25 44079 Tel 1 718 981 0401 Fax 39 ...