background image

CAEN

 

 

Electronic Instrumentation

 

 

UM2679 – A1536 AG536 Family Power Supply Boards

 

1.

 

Overview  

The Mod. A1536 – AG536 is a family of HV  boards, available with either positive or negative polarity, compatible 
with the CAEN Universal Multichannel Power Supply System (SY1527, SY2527, SY3527, SY4527, SY5527).  

The A1536 and A1536H channels share a common floating return, which allows on-detector grounding reducing the 
noise level; the floating return is insulated from the crate earth up to ±50 V (with a 65 V hardware limit);  the return 
of the AG536 channels is wired to the crate Earth reference. 

A1536 and AG536 output range is 0 ÷ 3 kV, with 1 mA maximum output current (800µA for A1536V and AG536V) 
and  100mV  set  and  monitor  resolution.  A1536H  output  channels  offer  dual  current  ranges  (software  selectable): 
High  Power:  0  ÷  1  mA  (Iset  res.:  20 nA;  Imon  res.:  1  nA;  1.5W  maximum  channel  power)  and  High  Resolution: 
0 ÷ 100  µA  (Iset  res.:  20nA;  Imon  res.:  100  pA);  it  features  also  50mV  and  5mV  set  and  monitor  resolution 
respectively.  

The boards are provided with both current and voltage protections. If overcurrent occurs, the relevant channel can 
be programmed either to turn off after a programmable trip time or to keep on providing the maximum allowed 
current: this particular feature allows the modules to work as current generator. The maximum output voltage can 
be fixed, through a potentiometer located on the front panel, at the same common value for all the board channels 
and this value can be read out via software.  

The HV RAMP-UP and RAMP-DOWN rates may be selected independently for each channel in the 1 ÷ 500 V/s range 
(1 V/s step).  

The 12 channel versions  output  voltages are  provided via  SHV connectors; 24 channel versions  are available with 
either SHV connectors or Radiall 52-pin connector; all other models feature a Radiall 52-pin connector. 24 channel 
SHV versions are double width boards (10 TE); all other versions are single width (5 TE).  “Multipin” versions have 
also the safety board interlock: this protection allows to disable the primary HV generation when the HV outputs 
are not connected to their loads. 

Tab. 1 – Available versions 

Version 

Ch Nr. 

Polarity 

Connector 

Return 

Size 

IMax 

A1536DN 

12 

Negative 

SHV 

Common Floating 

1U 

1 mA 

A1536DP 

12 

Positive 

SHV 

Common Floating 

1U 

1 mA 

A1536DM 

12 

Mixed 

SHV 

Common Floating 

1U 

1 mA 

A1536HDN 

12 

Negative, dual range 

SHV 

Common Floating 

1U 

1 mA 

A1536HDP 

12 

Positive, dual range 

SHV 

Common Floating 

1U 

1 mA 

A1536HDM 

12 

Mixed, dual range 

SHV 

Common Floating 

1U 

1 mA 

AG536DN 

12 

Negative 

SHV 

Ground 

1U 

1 mA 

AG536DP 

12 

Positive 

SHV 

Ground 

1U 

1 mA 

AG536DM 

12 

Mixed 

SHV 

Ground 

1U 

1 mA 

A1536LN 

24 

Negative 

Multipin 

Common Floating 

1U 

1 mA 

A1536LP 

24 

Positive 

Multipin 

Common Floating 

1U 

1 mA 

A1536HLN 

24 

Negative, dual range 

Multipin 

Common Floating 

1U 

1 mA 

A1536HLP 

24 

Positive, dual range 

Multipin 

Common Floating 

1U 

1 mA 

AG536LN 

24 

Negative 

Multipin 

Ground 

1U 

1 mA 

AG536LP 

24 

Positive 

Multipin 

Ground 

1U 

1 mA 

A1536SN 

24 

Negative 

SHV 

Common Floating 

2U 

1 mA 

A1536SP 

24 

Positive 

SHV 

Common Floating 

2U 

1 mA 

A1536SM 

24 

Mixed 

SHV 

Common Floating 

2U 

1 mA 

A1536HSN 

24 

Negative, dual range 

SHV 

Common Floating 

2U 

1 mA 

A1536HSP 

24 

Positive, dual range 

SHV 

Common Floating 

2U 

1 mA 

A1536HSM 

24 

Mixed, dual range 

SHV 

Common Floating 

2U 

1 mA 

AG536SN 

24 

Negative 

SHV 

Ground 

2U 

1 mA 

AG536SP 

24 

Positive 

SHV 

Ground 

2U 

1 mA 

AG536SM 

24 

Mixed 

SHV 

Ground 

2U 

1 mA 

A1536N 

32 

Negative 

Multipin 

Common Floating 

1U 

1 mA 

A1536P 

32 

Positive 

Multipin 

Common Floating 

1U 

1 mA 

A1536HN 

32 

Negative, dual range 

Multipin 

Common Floating 

1U 

1 mA 

A1536HP 

32 

Positive, dual range 

Multipin 

Common Floating 

1U 

1 mA 

AG536N 

32 

Negative 

Multipin 

Ground 

1U 

1 mA 

AG536P 

32 

Positive 

Multipin 

Ground 

1U 

1 mA 

A1536VN 

32 

Negative, dual range 

Multipin 

Common Floating 

1U 

800µA 

A1536VP 

32 

Positive, dual range 

Multipin 

Common Floating 

1U 

800µA

 

AG536VN 

32 

Negative 

Multipin 

Ground 

1U 

800µA

 

AG536VP 

32 

Positive 

Multipin 

Ground 

1U 

800µA

 

“Mixed” polarity means that half of the channels are negative and half positive 
 

Содержание A1536

Страница 1: ...CAEN Tools for Discovery n Electronic Instrumentation Rev 10 Es A1536 AG536 Family Power Supply Boards Rev 10 11 February 2015 ...

Страница 2: ...replace any product within the guarantee period if the Guarantor declares that the product is defective due to workmanship or materials and has not been caused by mishandling negligence on behalf of the User accident or any abnormal conditions or operations CAEN declines all responsibility for damages or injuries caused by an improper use of the Modules due to negligence on behalf of the User It i...

Страница 3: ...tion requirements 9 General safety information 9 Injury Precautions 9 Safety Terms and Symbols on the Product 9 Installation 9 3 Operating modes 10 Output control and monitoring 10 Output Enable 11 Grounding specifications 11 Safety Earth connection 11 List of Figures Fig 1 A1536 A1536D and A1536S front panel 6 Fig 2 JA jumper location 8 Fig 3 Earth configuration connection examples 12 List of Tab...

Страница 4: ...ons are single width 5 TE Multipin versions have also the safety board interlock this protection allows to disable the primary HV generation when the HV outputs are not connected to their loads Tab 1 Available versions Version Ch Nr Polarity Connector Return Size IMax A1536DN 12 Negative SHV Common Floating 1U 1 mA A1536DP 12 Positive SHV Common Floating 1U 1 mA A1536DM 12 Mixed SHV Common Floatin...

Страница 5: ...ec 1 Volt sec step Voltage Ripple 1 15 mV typical 30mV max 10 mV typical 15mV max VMon vs VOut Accuracy 2 3 typical 0 3 0 25 V VSet vs VOut Accuracy 2 typical 0 3 2 5 V IMon vs IOut Accuracy 2 typical 2 500 nA High Power typical 2 50 nA High Resolution typical 2 5 nA ISet vs IOut Accuracy 2 typical 2 1µA Maximum output power 3 W per channel 1 5W per channel Power consumption 160 W 120 W 60W full p...

Страница 6: ...CAEN Electronic Instrumentation UM2679 A1536 AG536 Family Power Supply Boards 6 Front Panel Fig 1 A1536 A1536D and A1536S front panel ...

Страница 7: ...nt Table 2 52 pin connector assignment 1 N C 11 Return 22 N C 32 Return 43 N C 2 N C 12 N C 23 N C 33 N C 44 N C 3 HVOUT23 13 N C 24 HVOUT12 34 HVOUT6 45 HVOUT0 4 HVOUT24 14 HVOUT18 25 HVOUT13 35 HVOUT7 46 HVOUT1 5 HVOUT25 15 HVOUT19 26 HVOUT14 36 HVOUT8 47 HVOUT2 6 HVOUT26 16 HVOUT20 27 HVOUT15 37 HVOUT9 48 HVOUT3 7 HVOUT27 17 HVOUT21 28 HVOUT16 38 HVOUT10 49 HVOUT4 8 HVOUT28 18 HVOUT22 29 HVOUT1...

Страница 8: ... 1 2 Agnd Shield Connects Agnd Earth to HV cable shield A1536 A1536L 3 4 Agnd Return Connects Agnd Earth to HV channels return A1536 A1536L A1536S 5 6 Shield Return Connects Shield to HV channels return A1536 A1536L On the 1536D RTN and AGND Earth may be connected by short circuiting JA jumper pins on the motherboard see figure below The protection shield must be screwed off in order to access JA ...

Страница 9: ...ut Covers To avoid electric shock or fire hazard do not operate this product with covers or panels removed Do Not Operate in Wet Damp Conditions To avoid electric shock do not operate this product in wet or damp conditions Do Not Operate in an Explosive Atmosphere To avoid injury or fire hazard do not operate this product in an explosive atmosphere Do Not Operate With Suspected Failures If you sus...

Страница 10: ...e OFF POn EN DIS Power On option which can be enabled or disabled If this option is enabled at Power On or after a Restart each channel is restored in the same condition defined by the Power parameter it was before the Power Off or Reset If this option is disabled at Power On or after a Restart all the channels are off independently from the condition in which they were before the Power Off or Res...

Страница 11: ...n 1 is the upper one Leave contact open between pin 3 and 4 The INTERLOCK LED red is turned off as one of the actions above is performed When the channels are disabled the voltage outputs drop to zero at the maximum rate available when the output disable cause is removed see above the channels remain OFF until the User turns them ON via software Grounding specifications The A1536 channels share a ...

Страница 12: ...ATION HV Generator Load PMT 0 HV Channel 0 RETURN Load PMT 31 HV Channel 31 EARTH of DETECTOR Cable SHIELD Circuit Protected SHIELD Circuit Protected RETURN HV CABLE AGND 2 1 4 3 6 5 FRONT PANEL JUMPERS CLOSED LOOP EARTH CONFIGURATION WITH PROTECTION STAGE ON LOAD GND HV Generator Load PMT 0 HV Channel 0 RETURN Load PMT 31 HV Channel 31 EARTH of DETECTOR Cable SHIELD Circuit Protected SHIELD Circu...

Страница 13: ...cluding engineers scientists and technical professionals who all trust them to help achieve their goals faster and more effectively CAEN S p A CAEN GmbH CAEN Technologies Inc Via Vetraia 11 Klingenstraße 108 1140 Bay Street Suite 2 C 55049 Viareggio D 42651 Solingen Germany Staten Island NY 10305 Italy Phone 49 0 212 254 4077 USA Tel 39 0584 388 398 Fax 49 0 212 25 44079 Tel 1 718 981 0401 Fax 39 ...

Отзывы: