ACCESSORIES SUPPLIES
DESCRIPTION
COMPLETE ASPIRATION JAR 1000cc
CONICAL FITTING
TUBES SET 6mm x 10mm (TRASPARENT SILICON)
ASPIRATION PROBE CH 20
ANTIBACTERIAL FILTER
The filter is produced with (PTFE) hydrophobic material witch prevents fluids entering the pneumatic circuit.
When the filter is wet, it’s not possible to use the unit therefore the filter should be changed immediately. In case of possible contamination or
discolouration, change the filter immediately.
Don’t use the suction unit without the protection filter fitted. If the suction unit is used in an emergency or in a patient where the risk of
contamination is not know the filter must be changed after each use. Available under request with different versions with complete jar 2000cc.
Suction catheter
: Single-use device to be used on a single patient. Do not wash or re-sterilize after use.
Reuse may cause cross-infections. Don’t use after lapse of the sell-by date
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 10993-1
standards on material biocompatibility.
PERIODICAL MAINTENANCE CHECKS
The
NEW ASKIR 20
suction equipment does not need maintenance or lubrication.
It is necessary to check functioning and instrument before every use. Unpack the instrument and
always check
integrity of plastic parts and feeding
cable, they might have been damaged during previous use. Connect cable to electrical network and turn switch on. Close the aspirator outlet with your
finger and with suction regulator in maximum vacuum position check that the vacuum indicators reaches -75kPa (-0.75 bar) minimum.
Rotate the knob from right to left and check the aspiration regulating control. The vacuum indicator should go down -25kPa (-0.25 bar). Verify that loud
noises are not present, these can indicate wrong functioning. A protection fuses
(F 1 x 1.6 A 250V)
reachable from exterior and situated in the plug
protects the instrument. For fuses replacing, always the type and the range.
Type of fault
Cause
Remedy
1. The suction unit doesn’t work
Cable is damaged
External power source failure
Replace the cable
Check the external power source
2. No aspiration
Jar Cap badly screwed down
Unscrewed the cap, then rescrew it correctly
3. No aspiration
Lid seal not in its seat
Unscrew the cap and insert the seal properly in its seat
4. The Vacuum power on the patient
side is either very low or absent
•
Vacuum regulator set to minimum
•
Protection filter blocked or damaged
•
Connection tubes blocked, kinked or
disconnected
•
Shut-off valve blocked or damaged
•
Pump motor damaged
•
Turn the vacuum regulator clockwise and check the
value of the vacuum on the gauge
•
Replace the filter
•
Replace or reconnect the tubes, check the jar
connections
•
Empty the jar, or disconnect the tube from the jar and
unblock the shut-off valve. The unit twill only work in
the upright position
•
Refer to authorised service personnel
5. The float doesn’t close
If the cap has been washed, ensure that
the float is not partially detached
Insert the float into it’s place
6. The float doesn’t close
The float it’s covered by dirty material
Unscrewed the cap, leave the and put in on autoclave
7. Low suction
Foam inside the jar
Fill the jar to 1/3 full of ordinary water
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7
None of the remedies has achieved the
desired results
Contact the seller or CA-MI After-sales Assistance
Service
If the overfill security system it’s activated, don’t proceede with the liquid aspiration.
If the overfill security system doesn’t work there are two cases:
1° case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological filter who avoid the liquid penetration inside the
device.
2° case – If both the security system doesn’t work, there is the possibility that liquid comes inside the device, in this case return the device to CA-MI
technical service.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT CA-MI
TECHNICAL SERVICE.
CA-MI DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO BE
TAMPERED.
13