Defect
Cause
Remedy
The Metal Mesh is
extremely stained.
The Metal Mesh is broken.
Replace the Medicine Cup with a new one.
3. Power button backlight is
on and after few second
turns off but no nebulisation
The drug chamber is not
inserted in the main unit or
it is not correctly inserted
Mount the drug chamber on the main unit and
verify it is correctly locked (limit switch
click)
4. Slow nebulisation
Very oily drug
Dilute the drug with a physiological solution
5. No nebulisation and
operation backlight of
power button is off
The power supply is the
wrong type or not
correctly inserted in the
socket
Verify the use of the original power supply
and that it is properly connected to a working
wall socket
Defects 1
-2-3-4-
5
None of the solutions are
effective
Contact your dealer or the Manufacturer
technical service
The Manufacturer declares that the information contained in this manual is consistent with the technical and safety
specifications of the device to which the manual refers. The stated technical specifications are updated to the date of
the publication of this document and only apply to the machine to which they are attached. The Manufacturer
reserves the right to make changes or improvements without prior notice in this document.
TECHNICAL SPECIFICATIONS
TYPE (93/42/EEC)
Class IIa Medical Device
MODEL
KIWI Plus
POWER
SUPPLY
100÷240V~ 50/60Hz with provided power supply (
mod.
HNEL050100WE or SC06W
-
0501000V
) or
3.7V
with internal rechargeable Lithium-Iones battery
2600mAh
POWER CONSUMPTION
2.0 W approx.
AUTONOMY WITH INTERNAL
BATTERY
approx. 220 minutes
(continuously) or 14 days with 2 daily
applications(1 appli
cation = 1 total nebulization of 2.5ml
NaCl0.9%)*
MAX RECHARGING TIME
Approx 4 Hours
USEFUL LIFE OF THE BATTERY
300 cycles of 100% full recharging
RESIDUAL VOLUME (2 spitting)
< 0.1ml
MMAD check production limit 100% of
devices (laser diffraction sys
tem) ****
< 4.5
µ
m
MINIMUM NEBULIZATION VOLUME
during 100% production checks ****
> 0.25 ml/min (2ml NaCl 0.9%)
MMAD (EN 13544-
1) **
1.51
µ
m
GSD (EN 13544-
1) ***
1.75
Содержание KIWI Plus
Страница 1: ...1...
Страница 6: ...6 KIWI Plus MESH PC 200 0 1 ml 93 MMAD 1 51 m CA MI 1 2 3...
Страница 7: ...7 4 5 6 KIWI Plus 7 8 9 ISO 10993 1 10 EN 60601 1 11 12 13 14...
Страница 8: ...8 15 93 42 EEC...
Страница 9: ...9 1 8 ml 8 cc 12 ml 8 ml...
Страница 10: ...10 2 4 L S...
Страница 12: ...12 3 2 100 220 28 1 1 2 5 ml NaCl 0 9 300 1 100 3 2 1 DC 4 100 1 300...
Страница 13: ...13 LED 4 20 90...
Страница 14: ...14 1 1 1...
Страница 15: ...15 2 1 2 5 3 30 4 Y 10...
Страница 17: ...17 2 180 5 2 1 0 3 2 3 4 5 1 2 3 4 5...
Страница 20: ...20 CA MI Srl 24 CA MI srl CA MI srl 93 42 BF II AC DC...
Страница 21: ...21 IPX2 1 2 15 2006 66 2012 19...
Страница 57: ...57...