EPS-300 

  3 

B. Introduction 

EPS-300 power supply is compact, space saving and light weight in design. It can be used 

with  many  electrophoresis  applications,  such  as  running  standard  PAGE  gels,  multiple  PAGE 

mini gels, horizontal/vertical agarose gels, semidry transfer and mini-gel tank transfer. The LED 

display reports power supply status. Parameters are set by pressing UP/DOWN arrow buttons in 

combination with a MODE button. Four pairs of output jacks accept 4 mm plugs, which operate 

in  parallel.  The  output  jacks  are  color  coded,  recessed  for  safety,  and  accept  both  4  mm 

unshielded  and  shielded  (banana)  plugs.  The  user  interface  consists  of  a  three  digit,  LED 

readout,  and  five  buttons  on  a  membrane  keypad.  Diagrams  of  the  front  and  rear  panel  are 

shown on the following pages. The EPS-300 operates in constant voltage or constant current 

modes.  Power  output  is  controlled  by  setting  a  maximum  value  for  voltage  (up  to  300  V),  or 

current (up to 500 mA). The maximum output power at any time is 90 Watts. The Run duration 

can  be  continuous  or  programmed  to  a  maximum  duration  of  1440  minutes.  The  EPS-300 

automatically  “crosses  over”  or  switches  the  limiting  parameter  according  to  the  programmed 

limits as the resistance of the electrical load changes during a run.   

 

B-1. Specifications 

Type 

  EPS-300 II ; EPS-300 IIV 

Input 

120V, 50-60 Hz ; 230V, 50-60 Hz 

Output voltage (V) 

10 – 300, 1 V steps 

Output current (mA) 

4 – 500mA, 1mA steps 

Maximum output power (W) 

90 W   

Output jacks 

4 pairs 

Output connector 

4-mm plugs 

Timer (Minutes) 

1 – 1440, 1 minute increment 

User interface 

3 digit, red LED membrane keypad 

Dimension (L x W x H) 

320 x 240 x 73 mm 

Weight 

Net ~2.4 kg (5.3 lb) Gross ~3.4 kg (7.5 lb) 

Safety features 

No load detection, Sudden load change protection,   

Over voltage protection, Automatic power failure recovery 

Operating conditions 

Indoor use: 0 – 40 

 

Relative humidity: 0% to 90% Non-condensing 

Altitude: < 2000 m 

Installation category: II 

 

 

 

Содержание EPS-300 II

Страница 1: ...EPS 300 II and EPS 300 IIV Power Supply Installation and Operation Manual Version 1 0 This instrument is intended for laboratory use only...

Страница 2: ...ct Description B 2 1 EPS 300 Power Supply Hardware Overview C Installation of EPS 300 Power Supply 5 C 1 Package List C 2 Setup and Operation of the EPS 300 Power Supply C 2 1 Unpacking C 2 2 Setup an...

Страница 3: ...anty EPS 300 is warranted to be free from defects in materials or workmanship for a period of one year from the original invoice date under normal usage Any defects occurring during warranty period C...

Страница 4: ...t the power supply using the AC power cord provided with the power supply or a similar 3 prong grounded AC outlet Use other than C B S Scientific s electrical cable with plugs or banana jacks is done...

Страница 5: ...utput is controlled by setting a maximum value for voltage up to 300 V or current up to 500 mA The maximum output power at any time is 90 Watts The Run duration can be continuous or programmed to a ma...

Страница 6: ...elected parameter 5 RUN Starts and Stops the output from the Power Supply The front and rear panel buttons as shown above are used to set up and run the EPS 300 power supply The audible alarm design a...

Страница 7: ...ive Inspect all the components for damage that may have occurred while the unit was in transit If any part appears damaged contact the carrier immediately Be sure to keep all packing material for dama...

Страница 8: ...un variable Press the MODE button to check and verify the set value of the other non constant run variable This variable is set by default to the maximum output of the power supply This is the normal...

Страница 9: ...clock symbol Press the UP and DOWN buttons to enter in the run time in minutes Press both the UP and DOWN buttons at the same time The display will briefly show PF0 indicating that the PFD mode is dis...

Страница 10: ...unction 4 Contact your local C B S Scientific representative Fan is spinning LED Display do not light up Front panel power on indicator off 1 Check main power source 2 Check if any errors on the front...

Страница 11: ...etected at some point after a successful start Buffer is leaking from the cells There is a loose electrical connection in the circuit 1 Check and correct all potential resistance problems 2 Check buff...

Страница 12: ...de Press any button except the RUN button to clear the error message D 3 Replacing the fuses 1 Unplug the power cord from the power supply 2 Locate the fuse compartment on the rear panel 3 Use a screw...

Страница 13: ...pply and allow to dry completely Please contact your local representative or C B S Scientific technical service specialist by E mail technicalservice cbssci com C B S Scientific offers complete techni...

Страница 14: ...z 5 200V 01 2A 200W max EPS 300 II Power Supply 120 V 50 60 Hz 10 300V 4 500mA 90W max EPS 300 IIV Power Supply 230 V 50 60 Hz 10 300V 4 500mA 90W max EPS 600 Power Supply 96 240 VAC 50 60 Hz 10 600V...

Страница 15: ...C B S Scientific Company P O Box 856 Del Mar CA 92014 Phone 858 755 4959 email technicalservice cbssci com www cbsscientific com Printed in Taiwan...

Отзывы: