EPS-300
10
E7
Short Circuit
Load current exceeded 500 mA
1. Check for any short circuit or evidence
of electrical arching.
2. Buffer concentration is too high.
3. Press any button* to clear the error
code.
*Press any button, except the RUN button, to clear the error message.
D-3. Replacing the fuses
1.
Unplug the power cord from the power supply.
2.
Locate the fuse compartment on the rear panel.
3.
Use a screwdriver to pry out the fuse compartment. There are two fuses in the compartment.
Check and replace any blown fuses with a fuse of the same size and ratings.
4.
FUSES (2); 250 V, 3.15 A, 5 x 20 mm fast blow.