GasAlertMicro
Users Manual
Language Selection
The detector is shipped with English as the displayed
language. Multiple languages are available. To choose the
desired language, press
H
or
G
. Press
C
to select the new
language.
Sensor Enable or Disable
a
Warning
Disabling an installed sensor configures the detector
to a 1, 2, or 3-gas unit. No protection is now provided
for the gas targeted by that sensor(s).
Note:
Disabling a sensor should be performed with extreme
caution.
In the event a sensor fails, Sensor Disable can be used to
turn off the sensor fail alarm. The sensor should be replaced
and enabled as soon as possible.
The detector will function normally with remaining enabled
sensors. The sensor may be enabled again, at any time.
To disable (or enable) the H
2
S
sensor, press
C
when the display
advises:
The display then advises the H
2
S
sensor is
OFF
.
To enable the H
2
S sensor, repeat
the above sequence. The display
advises H
2
S sensor is
On
.
Repeat the sequence above for other sensors.
Note:
If all the sensors are turned off the display will advise:
All sensors are disabled - all sensors
are reset to on
.
Note:
If a sensor is turned on through the display and the
sensor is not installed in the detector the display will advise:
H2S sensor not installed - sensor cannot
be enabled
16