ProDAQ 3150 User Manual
3150UM-01
Bustec Production Ltd.
Page 45
TRIG_IN_EN_IRQ (0x00C0 0018, read/write)
Bit
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Operation RW RW RW RW RW RW RW RW RW RW RW RW RW RW RW RW
Initial
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
Name
VXITI
7
VXITI
6
VXITI
5
VXITI
4
VXITI
3
VXITI
2
VXITI
1
VXITI
0
FCTO
7
FCTO
6
FCTO
5
FCTO
4
FCTO
3
FCTO
2
FCTO
1
FCTO
0
FCTO 0 .. 7
Selects the FC trigger signals, which will generate a trigger interrupt for the DSP
(IRQ2)
VXITI 0 .. 7
Selects VXI trigger Signals, which will generate a trigger interrupt for the DSP
(IRQ2)
A write operation with a one to the appropriated bit enables the trigger generation.
In case of a read the bits which have been enabled and have caused an interrupt are active.
The bits which have caused an interrupt have to be cleared by disabling the appropriated bit.
TRIG_OUT (0x00C0 001A, read/write)
Bit
15
14
13 12 11 10
9
8
7
6
5
4
3
2
1
0
Operation
RW
RW
RW RW RW RW RW RW RW RW RW RW RW RW RW RW
Initial
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
Name
LIST_FLAG2
_IRQ2
FL2 FL1 FL0 TMO
2
TMO
1
TMO
0
FCTI
7
FCTI
6
FCTI
5
FCTI
4
FCTI
3
FCTI
2
FCTI
1
FCTI
0
Read:
FCTI 0 … 7
Status of the Function card trigger input (actual level; “0” : active; “1” : not active)
TMO 0 … 2
Status of the trigger line from the communication registers to the trigger matrix
(actual level; “0” : active; “1” : not active)
FL0 … FL2
“0”
LIST_FLAG2_IRQ2 Status of the DSP IRQ2 LIST_FLAG2 flag
Write:
FCTI 0 … 7
“1” : set the Function card trigger input (set to low)
“0”: clear the Function card trigger input (tristate)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com