Megohmmeter
Model 24508
Page 13
Technical data
General informa
-
tion
Pr
eparations for
use
Contr
ols and con
-
nections
Manual operation
of the unit
Contr
olling the
unit r
emotely
Maintenance and customer service
2. Preparations for use
Pr
eparations for
use
2.1 Unpacking the unit
The unit weighs 2.1 kg and is packaged accordingly to protect against
shock.
Carefully unpack the unit and verify that all items are present.
This normally includes: 1 digital megohmmeter type 24508
1
power
lead
1 pair of instrument leads
1 copy of this manual.
Inspect the unit carefully for damage.
If you suspect that the unit has been damaged during shipping, notify the de-
livery company within 72 hours.
The packaging should be retained for examination by a representative of the
manufacturer and/or the delivery company.
The megohmmeter type 24508 should be shipped only in its original packaging
or in packaging capable of providing an equivalent degree of protection.
2.2 Using the equipment for the first time
Check that the correct supply voltage 230 V/50 Hz is available. Connect the
unit to a standard grounding outlet using the power lead supplied.
Warning:
The unit must not be switched on under any circumstances if it
shows signs of damage during shipping.
The case or measurement input can carry life-threatening voltages if the mains
voltage is transferred as a result of damage.
2.3 Supply voltage
The supply voltage is factory set to 230 V~.
Supply voltage:
230 V~ ± 10 %
Frequency range:
50 Hz / 60 Hz
Power consumption:
10 VA
Fuse rating:
0.16 AF