8.2
installing a device on the pipe
Danger
risk of injury due to high pressure in the installation.
▶
Before any intervention in the installation, stop the circulation of fluid, cut off the pressure and drain the pipe.
▶
Before any intervention in the installation, make sure there is no pressure in the pipe.
risk of injury due to the nature of the fluid.
▶
Respect the prevailing regulations on accident prevention and safety relating to the use of dangerous fluids.
If a pH/redox ("ORP") probe (with PG 13.5 head, 120 mm long and without temperature probe) from a
supplier other than Bürkert is used, follow the relevant instructions on installation in the pipe.
Direction of the fluid
→
Choose an appropriate position in the pipe to install
the fitting.
→
Fit the pipe with a fitting with G 1” ½ external
threaded sensor connection with respect to the
instructions delivered with the fitting.
Fig. 12: Mounting positions of the device in the pipe
75°
75°
→
Fit the fitting at an angle of ±75° max. to the vertical
in order to ensure the good operation of the pH/redox
("ORP") probe.
Fig. 13: Angle to the vertical
The probe must always be immersed in the fluid to prevent it drying out.
→
Fit the display module. Refer to chpt. 7.4.
→
Calibrate the device. Refer to chpt. 9.12.