SAFETY INSTRUCTIONS
•
Always consult the patient’s physician before using the Burke Low Air Loss Support
System for contra indicated conditions.
•
To avoid damaging your Burke Low Air Loss Support System control unit (A), before
operating be sure the AC power (X) available at your location matches the power require-
ments printed on the product identification label on the back of the unit.
•
To avoid electric shock, always plug in the power cord of the control unit into a prop-
erly grounded power source (X).
•
Do not insert items into any openings of the control unit (A). Doing so may cause
fire or electrical shock by shorting internal components.
•
Do not spill liquids or food on or into the control unit (A). In the event of any spill-
age, immediately turn off the control unit and disconnect it form the power source (X). Re-
turn the control unit for servicing to a factory authorized service center.
•
Care should be taken such that the controls on the footboard of the bed frames are
not obstructed by the Burke Low Air Loss Support System control unit.
•
Care should be taken such that the control unit (A) is not blocked, and kept away
from any heat sources or radiators during the operation of the unit.
•
Care should be taken such that the power cord (Q) of the control unit is not pinched,
or has any objects placed on it, and also ensure it is not located where it can be stepped on
or tripped over.
•
Do not attempt to service the control unit except as explained in this operating
instructions manual. Contact factory for servicing instructions. Always follow operating and
service instructions closely.
•
Do not place the patient directly on the mattress without the top sheet. Breathable
nylon top sheet is water repellent, highly vapor permeable, anti-microbial, low friction and
low shear quilted reusable top sheet.
•
The control unit enclosure should only be opened by a factory authorized quali-
fied service technician.
•
Smoking by the patient or anyone else around or on the Burke Low Air Loss Sup-
port System is prohibited. Burke Low Air Loss Support System uses room air for circulation
through the mattress, smoking will contaminate the system.
!
!
!
!
!
!
!
!
!