background image

PA30 Series Powered Air-Purifying Respirator

PA3/PA3IS Blower Assembly User Manual

7

Mounting and Replacing Cartridges on the
Blower Unit

The useful life of a chemical cartridge for vapors and gases will
vary with the concentration and nature of the contaminant, the
breathing rate of the respirator wearer, and ambient temperature
and humidity.

The Occupational Safety and Health Administration (OSHA)
regulations 29 CFR 1910.134 require that the employer must
implement a change schedule when using Air-Purifying respirators
for protection against gases and vapors, if there is no end-of-
service-life indicator on the cartridges. The change schedule must
be based on objective data that will ensure that the cartridges are
replaced before the end of their service life. Factors to consider
include workplace conditions such as contaminant concentration,
relative humidity, temperature, work activities, respirator use
pattern (e.g., continuous or intermittent use), presence of other
contaminants, potential for contaminant migration/desorption,
health effects of the gas or vapor, and the presence of any
warning properties. Contact Bullard for further information on
change-out schedules.

High efficiency particulate filters must be replaced when retained
particles clog the filters and reduce air flow below acceptable
levels, as indicated by testing with the Air Flow Indicator. 
See page 5. 

To Replace Cartridges

Remove the air-purifying element from its packaging, and inspect
for damage. If in doubt do not use.

Check that the air-purifying element has not exceeded its “use-by”
date and that the connecting thread is in good condition.

Check that the air-purifying element is appropriate to the hazard.
If in doubt consult your respirator program administrator or
supervisor.

Check that the threads in the blower unit port are in good
condition and clear of contaminant.

Check that the PA3BU/PA3ISBU blower ports have the rubber
gasket seals present.

Remove the air-purifying element seals 

(see Figure 4)

.

Screw the air-purifying elements into the receptacles 

(see

Figure 5)

until the cartridge is hand tight. 

DO NOT OVER-

TIGHTEN. 

When using the two PA3HE filter cartridges, install the
filter plug into one of the ports. DO NOT OVER-TIGHTEN
(Figure 6).

Mounting the Cartridges

Figure 5

Figure 4

www.bullard.com

Figure 6

Содержание PA30 Series

Страница 1: ...espiratory protection during abrasive blasting or clean up 4 Do not use in circumstances where the airborne concentration level of contaminant exceeds maximum use concentration for this type of respirator as established by regulatory standards 5 Leave area immediately if Breathing becomes difficult Dizziness or other distress occurs You taste or smell the contaminant Unit becomes damaged Voltage a...

Страница 2: ...erational Inspection 4 Mounting the Breathing Tube 5 Checking Air Flow 5 Air Purifying Elements 6 Mounting and Replacing Cartridges 7 Donning the Blower 8 Low Battery Alarm 8 Troubleshooting 9 Cleaning and Storage 9 NIOSH Approval Label 10 11 Ordering Information 12 Table of Contents 2 www bullard com ...

Страница 3: ...Hat 20SICVH Hood with Taped and Sealed Seams and CBRN resistant PVC lens for use with Bullard Hard Hat 20TPC Hood with Solvent Resistant Lens and Inner Bib 20TP Hood with Solvent Resistant Lens 20LFM Loose Fitting Facepiece Hood Medium Size 20LFL Loose Fitting Facepiece Hood Large Size 6 The Battery Charger PA3SMC Quick Charger PA3GC Gang Charger PA3ISC Intrisically Safe Quick Charger PA3ISGC Intr...

Страница 4: ... is fully charged This procedure should also be followed after periods of prolonged storage Without periodic charging the nickel metal hydride batteries will lose up to 1 2 of their charge per day Allowing a battery to self discharge during periods of prolonged storage will not damage the battery Pre Operational Inspection Prior to each work shift perform the following Pre Operational Inspection t...

Страница 5: ... not blocked Two hands may be used if preferred one to hold the breathing tube and one to hold the Airflow Indicator The position of the ball in the Airflow Indicator should be observed If any part of the ball is below the PASS LINE on the Airflow Indicator check for Blower malfunction Clogged or damaged Air Purifying filter elements on the HE filters or the combination cartridges with HE filters ...

Страница 6: ...inst the following contaminants independent laboratories have tested and verified that the PA3NBC cartridge is effective against many chemical warfare agents and or their recognized surrogates These agents surrogates are listed below WARNING Use only the cartridge described in the above table Used particulate laden cartridges must be changed as a set and not individually All cartridges must be of ...

Страница 7: ...cts of the gas or vapor and the presence of any warning properties Contact Bullard for further information on change out schedules High efficiency particulate filters must be replaced when retained particles clog the filters and reduce air flow below acceptable levels as indicated by testing with the Air Flow Indicator See page 5 To Replace Cartridges Remove the air purifying element from its pack...

Страница 8: ... if the battery voltage is below acceptable levels The PA3BU Low Voltage Alarm is an electronic 77 dba beep The PA3ISBU Low Voltage Alarm is a mechanical pulsing of the blower In either case the alarm is internal to the blower and the sound will be carried up the breathing tube into the hood Sounding of the alarm indicates that insufficient airflow may be imminent The user should immediately do th...

Страница 9: ...ne organic based solvents or chlorinated degreasing fluids such as trichloroethylene as they will cause damage DO NOT immerse the equipment in water or other cleaning fluid as this may cause contamination in the breathing tube and blower interior that will be difficult to remove Failure to observe the instructions and warnings in this manual invalidates all performance statements and approvals for...

Страница 10: ...s for changing cartridges and or filters M All approved respirators shall be selected fitted used and maintained in accordance with MSHA OSHA and other applicable regulations N Never substitute modify add or omit parts Use only exact replacement parts in the configuration as specified by the manufacturer O Refer to users instructions and or maintenance manuals for information on use and maintenanc...

Страница 11: ... for changing cartridges and or filters M All approved respirators shall be selected fitted used and maintained in accordance with MSHA OSHA and other applicable regulations N Never substitute modify add or omit parts Use only exact replacement parts in the configuration as specified by the manufacturer O Refer to users instructions and or maintenance manuals for information on use and maintenance...

Страница 12: ...olvent resistant polyester lens no suspension CC20 only Hood with inner bib for use with Bullard hard hat 20TICH Tychem QC hard hat not included 20SICH Tychem SL with taped and sealed seams hard hat not included 20SICVH Tychem SL with taped and sealed seams and CBRN resistant PVC lens hard hat not included Hood with inner bib and long outer bib for use with headband suspension 20TICS Tychem QC wit...

Отзывы: