EGK 4S
Connection via plug
The device is equipped with connectors according to EN 175301-803 for mains and status output. If the cables are mounted prop-
erly, they cannot be interchanged. Please make sure that the connectors are remounted correctly after connecting the cable. The
following figures show the pin assignment with respect to the numbers printed on the connector.
The supply line cross-sections must be suitable for the rated current. Use a maximum line cross-section of 1.5 mm² (AWG 16) and
a cable diameter of 8 - 10 mm (0.31 - 0.39 inch).
power supply
status output
pin assignment
function ok
alarm
Fig. 1: A05-100002 Cooler electric supply
The mains supply must be protected with 10 A. The clamping area has a diameter of 8-10 mm. If the unit features a peristaltic
pump, it must be connected to a power source separately.