background image

GAS 222.21 AMEX

WARNING

Gas emanation

Sample gas can be harmful to the health!

Check the lines for leaks.

4.7.1 Blowback Connection

If the probe has a G3/8 blowback connection, without factory
installed blowback device it will be open. The blowback con-
nection must be sealed gas-tight prior to start-up.

DANGER

Toxic, corrosive gasses

Explosive or toxic gases can develop due to a leaking or open
blowback connection.

4.7.2 Connecting the Gas Line

Please note the following items when connecting the sample
gas line (NPT 1/4") on heated probes to prevent thermal
bridges:

• Choose the shortest possible screw connection.
• Shorten the connection pipe for the sample gas line as

much as possible. To do so, remove the insulation around
the sample gas line. This is done by loosening the fixing
bolts.

CAUTION

Fragile

The insulation is fragile. Handle with care, do not drop.

After connecting the sample gas line it must be braced and se-
cured with the clamp.

Long sample gas lines may require additional support clamps
along the way to the analysis system! Once all lines have been
connected and checked for leaks, carefully reinstall and secure
the insulation.

WARNING

Gas emanation

Sample gas can be harmful to the health!

Check the lines for leaks.

4.7.3 Connecting the calibrating gas line (optional)

Connecting the calibrating gas line requires a Ø6 mm or Ø1/4”
pipe fitting.

If the calibrating gas connection was ordered with check
valve, a Ø6 mm or Ø1/4” pipe can be connected directly to the
check valve.

4.8 Connecting the Blowback and Pressure Vessel
(Optional)

The air lines must be connected carefully and properly, using
suitable fittings.

If the probe is equipped with pressure vessel for efficient
blowback (optional), a manual shut-off valve (ball valve) must
be installed in the air supply, immediately upstream from the
pressure vessel.

On probes used to sample flammable gas, nitrogen (inert gas)
must be used for blowback. Blowback of explosive gases is
prohibited.

NOTICE

The operating pressure of the compressed air (inert gas) re-
quired for blowback must always be higher than the process
pressure.

Required pressure differential min. 3 bar (44 psi).

DANGER

Broken pressure vessel

Gas leak, danger due to flying parts.

Maximum operating pressure of the pressure vessel 10 bar
(145 psi)!

The operating pressure reduces based on the operating
voltage (see solenoid valve type plate).

DANGER

Adiabatic compression during gas blowback (explosion
hazard)!

Adiabatic compression may cause high gas temperatures and
must be checked by the user.

Gas blowback may result in high gas temperatures due to
adiabatic compression. This can cause flammable gases to ig-
nite spontaneously.

a) Blowback of explosive atmosphere / gases is prohibited.

b) Flammable atmosphere / gases (non-explosive) may only

be blown back with nitrogen (inert gas).

4.9 Electrical connections

WARNING

Hazardous electrical voltage

The device must be installed by trained staff only.

CAUTION

Wrong mains voltage

Wrong mains voltage may damage the device.

Regard the correct mains voltage as given on the type plate.

CAUTION

Equipment damage

Cables damaged

Do not damage the cable during installation. Install a strain
relief for the cable connection. Secure the cable against twist-
ing and loosening. Please note the temperature resistance of
the cables (> 100 °C/212 °F).

The probe includes two cubic plugs per EN 175301-803 (ex
DIN43650) and a junction box. The plugs are prewired to the
junction box. For safety reasons this connection must not be
modified. The electrical connections must only be run to the
terminals of the junction box.

The power supply for the two heating cartridges (mains sup-
ply 115/230V, 50/60 Hz) and the connection for the thermal
alarm contact are provided via the terminal strip in the junc-

11

Bühler Technologies GmbH

BX460050 ◦ 01/2020

Содержание GAS 222.21 AMEX

Страница 1: ...460050 Version 01 2020 Ref 998BX460050 Analysentechnik Kurzanleitung Sonden deutsch 2 Brief Instructions Probes english 8 Notice de montage Sondes fran ais 14 Gu a r pida Sondas espa ol 20 chinese sim...

Страница 2: ...en aufgrund m glicher adiaba tischer Kompression hoher R cksp ldruck gegen verschmutz ten Filter nicht zul ssig Das Einhalten dieser Bedingungen liegt im Verantwortungsbereich des Betreibers unter Zuh...

Страница 3: ...osta tischen Entladungen kommen kann Reinigen Sie Geh useteile aus Kunststoff und Aufkleber nur mit einem feuchten Tuch Funkenbildung Sch tzen Sie das Betriebsmittel vor externen Schlageinwir kungen F...

Страница 4: ...en Bringen Sie elektrische Betriebsmittel die zur Instandhaltung ge ffnet werden m ssen m glichst in einen staubfreien Raum Falls dies nicht m glich ist verhindern Sie das Eindrin gen von Staub in das...

Страница 5: ...ann zerbrechen Vorsichtig behandeln nicht fallen lassen Nach Anschluss der Messgasleitung ist diese durch die Schelle abzufangen und zu sichern Bei l ngeren Messgasleitungen sind unter Umst nden weite...

Страница 6: ...aturbest n digkeit 100 C besitzen Achten Sie auf ausreichende Zugentlastung des An schlusskabels Kabeldurchmesser dem Dichtungsring der W rfelstecker anpassen Beachten Sie dass beim Heizsystem kurzzei...

Страница 7: ...t sind keine Teile der Gasentnahmesonde demontiert sind die Schutz und berwachungsvorrichtungen vorhanden und funktionsf hig sind z B Flammensperre der Gasein und Gasausgang der Gasentnahmesonde nicht...

Страница 8: ...he probes is prohibited due to possible adiabatic com pression high blowback pressure against contaminated fil ter The operator is responsible for compliance with these conditions taking into account...

Страница 9: ...tions do not frequently produce flammable electro static discharge Always clean plastic housing parts and decals with a damp cloth Sparking Protect the equipment from external blows Flame propagation...

Страница 10: ...amage to the system If the process holds a risk of flame propagation install a flame arrestor 4 3 Installing the sampling tube optional The sampling tube if necessary with the fitting extension must b...

Страница 11: ...pstream from the pressure vessel On probes used to sample flammable gas nitrogen inert gas must be used for blowback Blowback of explosive gases is prohibited NOTICE The operating pressure of the comp...

Страница 12: ...g installation Install a strain relief for the cable connection Secure the cable against twist ing and loosening Please note the temperature resistance of the cables 100 C 212 F DANGER Adiabatic compr...

Страница 13: ...tor s manual on the included CD or online at www buehler technologies com for maintenance informa tion 7 Service and Repair Please refer to the original operator s manual on the included CD or online...

Страница 14: ...la LIE et la LSE avec les sondes n est pas autoris en raison d une possible compression adiabatique pression de r trolavage lev e contre le filtre encrass Le respect de ces conditions est de la respo...

Страница 15: ...h tiques du bo tier ainsi que les au tocollants uniquement avec un linge humide Formation d tincelles Prot gez le moyen de production des chocs externes Retour de flammes En cas de risque de retour de...

Страница 16: ...a temp rature maximum de surface de la sonde respectez les normes et la r glementa tion en vigueur Placer si possible l quipement de production lectrique de vant tre ouvert des fins d entretien dans u...

Страница 17: ...a le re v tement isolant voire les mors isolants dans la zone de la conduite de gaz de mesure Desserrez pour cela les vis de fixation ATTENTION Risque de rupture Le mat riau isolant peut se briser man...

Страница 18: ...ment Le raccor dement de la sortie d alarme doit tre fait par le biais d un cir cuit limit en nergie selon EN 60079 14 et 15 Umax 30 V Imax 100 mA Le raccordement se fait selon le plan d implanta tion...

Страница 19: ...ifiez que qu les raccordements des tuyaux lectriques ainsi que le ru ban chauffant ne sont pas endommag s et qu ils sont cor rectement mont s aucun l ment de la sonde de pr l vement de gaz n est d mon...

Страница 20: ...LSE con las sondas debido a la posible compresi n adiab tica alta presi n de retrolavado contra el filtro sucio El cumpli miento de estas condiciones es responsabilidad del operador con ayuda de su e...

Страница 21: ...hispas Los equipos el ctricos solo deben instalarse en lugares en los que en un uso normal no suelan aparecer cargas electroest ticas inflamables Limpie las partes de pl stico de la carcasa y los adhe...

Страница 22: ...perfi cie del aparato tenga en cuenta las normas y regulaciones le gales vigentes Si es posible lleve los equipos el ctricos que deban abrirse pa ra su mantenimiento a un espacio libre de polvo Si no...

Страница 23: ...capa aislante y quitar tambi n las mordazas aislantes de la zo na del conducto de gas de muestreo Esto se lleva a cabo soltando los tornillos de fijaci n CUIDADO Peligro de rotura El material aislant...

Страница 24: ...onexi n se reali za de acuerdo al esquema de asignaci n adjunto Para conectar el suministro el ctrico utilice nicamente cables con una resistencia t rmica de 100 C Aseg rese de que el cable de conexi...

Страница 25: ...o est desmontada Los dispositivos de protecci n y control necesarios est n disponibles y operativos p ej supresor de llamas Las salidas y entradas de gas de la sonda de gas de mues treo no est n bloqu...

Страница 26: ...49 0 21 02 49 89 20 1 1 ID 25 1 2 GAS 222 21 AMEX 100105177 462222116633000999999 001 CI 1 Div 2 Gps B C D T3 CSA Certificate 1728394 Install per drawing 46 115 Z01 06 4 115 230 V max 360W 230V at 20...

Страница 27: ...GAS 222 21 AMEX 3 20 C 50 C 4 F bis 122 F 4 4 1 4 2 4 3 4 4 O O O 90 4 5 4 6 27 B hler Technologies GmbH BX460050 01 2020...

Страница 28: ...NPT 1 4 G1 8 G1 4 NPT 1 4 1 1 2 GAS 222 xx ANSI GAS 222 xx AMEX 4 7 1 G3 8 4 7 2 NPT 1 4 4 7 3 6 mm 1 4 6mm 1 4 4 8 3 bar 44 psi 10 bar 145 psi a b 4 9 100 C 212 F EN 175301 803 DIN43650 115 230V 50...

Страница 29: ...ax 30 V Imax 100 mA 1 kV 1 5 kV 4 9 1 PTC 115 230 V AC 100 C 100 C 212 F a b 5 5 1 Umax 30 V Imax 100 mA 6 CD www buehler technologies com 7 CD www buehler technologies com 29 B hler Technologies GmbH...

Страница 30: ...nd 49 0 21 02 49 89 0 49 0 21 02 49 89 20 1 1 o 25 1 2 Year 2019 CSA GAS 222 21 AMEX 100105177 462222116633000999999 001 Cl 1 Div 2 Gps B C D T3 CSA Certificate 1728394 Install per drawing 46 115 Z01...

Страница 31: ...GAS 222 21 AMEX a b c d a b c a b c 3 20 C 50 C 4 F 122 F 31 B hler Technologies GmbH BX460050 01 2020...

Страница 32: ...GAS 222 21 AMEX 4 4 1 4 2 4 3 4 4 90 4 5 4 6 32 B hler Technologies GmbH BX460050 01 2020...

Страница 33: ...1 3 2 1 DN65 PN6 DN3 150 2 G3 4 NPT 1 4 G3 8 1 6 1 4 2 NPT 1 4 G1 2 NPT 1 4 G1 8 G1 4 NPT 1 4 1 1 2 GAS 222 xx ANSI GAS 222 xx AMEX 4 7 1 G3 8 4 7 2 NPT 1 4 4 7 3 6 1 4 6 1 4 4 8 3 44 psi 33 B hler Te...

Страница 34: ...2 21 AMEX 10 145 psi a b 4 9 100 C 212 F 4 EN 175301 803 ex DIN43650 115 230 50 60 EN 60079 14 15 U 30 I 100 100 C 4 6 8 U 30 I 100 1 1 5 4 9 1 115 230 AC 100 C 34 B hler Technologies GmbH BX460050 01...

Страница 35: ...GAS 222 21 AMEX 100 C 212 F a b 5 5 1 U 30 I 100 6 www buehler technologies com 7 www buehler technologies com 35 B hler Technologies GmbH BX460050 01 2020...

Страница 36: ...power supply Customer alarm output Probe power supply Heater power supply PAV 01 Alarm output M16 x 1 5 mm 4 5 10 mm M16 x 1 5 M16 x 1 5 2 Connection diagram heated pressure vessel Heater Operating vo...

Отзывы: