If a large amount of condensate accumulates, we recommend
using a condensate trap with automatic condensate drain.
Our condensate drains, 11 LD spec., AK 20 V, or model 165 SS,
are suitable.

Glass vessels and automatic condensate drains are available
for draining condensate for external mounting below the
unit. When using automatic condensate drains, the sample
gas pump must be installed upstream of the cooler (pressure
operation) to ensure proper function of the condensate drain.

If the sample gas pump is located at the cooler outlet (suction
operation), we recommend using glass condensate traps or
peristaltic pumps.

Connecting the condensate drains

Depending on the material, build a connecting line with fit-
tings and tubing or hose between the heat exchanger and
condensate drain. For stainless steel the condensate drain can
be suspended directly to the connecting tube, for hoses the
condensate drain must be secured separately using a clamp.

The condensate drain can be mounted directly to the heat ex-
changer.

Condensate lines must always be installed with a slope and a
minimum inside diameter of DN 8/10 (5/16”).

4.2.1 Peristaltic pump connector (optional)

Coolers ordered with attached peristaltic pump already have
it installed and wired. Heat exchangers ordered at the same
time are already installed and connected to the peristaltic
pump.

The ø6 mm (0.24 inch) hose nipple for the pump’s condensate
outlet must be carefully and properly connected with a suit-
able hose and hose clamp.

Versions with screw connections DN 4/6 or 1/6”-1/4” are sup-
plied with ferrule and knurled nut and must be carefully
sealed with appropriate hose.

NOTICE

Installing peristaltic pumps CPsingle / CPdouble limits the
maximum permissible operating pressure in the system!

Operating pressure ≤ 1 bar

4.3 Electrical connections

The operator must install an external separator for the device
which is clearly assigned to this device.

This separator

• must be located near the device,
• must be easy for the operator to reach,
• must comply with IEC 60947-1 and IEC 60947-3,
• must separate all live conductors and the status output,

and

• must not be attached to the power feed.

WARNING

Hazardous electrical voltage

The device must be installed by trained staff only.

CAUTION

Wrong mains voltage

Wrong mains voltage may damage the device.

Regard the correct mains voltage as given on the type plate.

WARNING

High voltage

Damage to the device in case of insulation testing

Do not proceed insulation tests with high voltage to the
device as a whole!

Insulation test

The device is equipped with extensive EMC protection. If insu-
lation tests are carried out the electronic filter devices will be
damaged. All necessary tests have been carried out for all con-
cerned groups of components at the factory (test voltage 1 kV
or 1.5 kV respectively, depending on the device).

If you wish to carry out the insulation test by yourself, please
test only separate groups of components.

Disconnect the compressor, the fan, the heating or the peri-
staltic pumps, respectively, and then carry out the insulation
tests.

Connection via plug

The device is equipped with connectors according to EN
175301-803 for mains and status output. If the cables are
mounted properly, they cannot be interchanged. Please make
sure that the connectors are remounted correctly after con-
necting the cable. The following figures show the pin assign-
ment with respect to the numbers printed on the connector.

The supply line cross-sections must be suitable for the rated
current. Use a maximum line cross-section of 1.5 mm²
(AWG 16) and a cable diameter of 8 - 10 mm (0.31 - 0.39 inch).

pin assignment

power supply

Fig. 1: 

A100048 Cooler electric supply

The supply voltage is 230 VAC 50/60 Hz or 115 VAC 50/60 Hz
(please note type plate!). The mains supply must be protected
with 10 A. The clamping area has a diameter of 8-10 mm.

4.4 Signal outputs

At the back of the unit is a 12-pin PHÖNIX connector which
provides various status signals. The maximum switching
power of the alarm outputs is 250 VAC/DC, 1 A each.

An alarm is triggered if the temperature of the cooler is out-
side the specified limits. It does not indicate if the alarm was
triggered due to excess temperature or insufficient temperat-
ure.

8

Bühler Technologies GmbH

BX450010 ◦ 10/2018

Содержание EGK 2-19

Страница 1: ...urzanleitung Kompressor Messgask hler deutsch 2 Brief Instructions Compressor sample gas cooler english 7 Notice de montage Refroidisseur de gaz de mesure de compresseur fran ais 12 Gu a r pida Compre...

Страница 2: ...en eingehal ten werden Schutzeinrichtungen verwendet werden und vorgeschrie bene Wartungsarbeiten durchgef hrt werden bei der Entsorgung die gesetzlichen Regelungen beachtet werden GEFAHR Elektrische...

Страница 3: ...f ltig mit passendem Schlauch zu verbinden HINWEIS Durch den Einbau von peristaltischen Pumpen CPsingle CP double wird der maximal zul ssige Betriebsdruck im System eingeschr nkt Betriebsdruck 1 bar 4...

Страница 4: ...ratur D Masse des Ger ts An schluss der Schirmung der 4 20 mA Signallei tung 5 Betrieb und Bedienung HINWEIS Das Ger t darf nicht au erhalb seiner Spezifikation betrieben werden Nach dem Einschalten d...

Страница 5: ...ptmen Einstellung der K hlertemperatur im Bereich 2 C 20 C 35 6 F 68 F Setzen der oberen Alarmgrenze im Bereich tEMP 1 C 7 C 1 8 F 12 6 F Setzen der unteren Alarmgrenze im Bereich tEMP 1 C 3 C 1 8 F 5...

Страница 6: ...einhalten Verminderter Gasdurchsatz Gaswege verstopft W rmetauscher demontieren und reinigen ggf Filterelement austauschen Kondensataus gang vereist K hler einsenden Fehlermeldungen im Display Die An...

Страница 7: ...tory maintenance is performed Legal regulations are observed during disposal DANGER Electrical voltage Electrocution hazard a Disconnect the device from power supply b Make sure that the equipment can...

Страница 8: ...1 and IEC 60947 3 must separate all live conductors and the status output and must not be attached to the power feed WARNING Hazardous electrical voltage The device must be installed by trained staff...

Страница 9: ...lock temperature has reached the preset target value adjustable alarm range The status contact is in the Alarm position Once the target temperature range has been reached the temperature will continuo...

Страница 10: ...libration Exit Submenu C F Global Settings Exit Main Menu Adjust the set point 2 C 20 C 35 6 F 68 F Set the upper alarm threshold above setpoint tEMP 1 C 7 C 1 8 F 12 6 F Set the lower alarm threshold...

Страница 11: ...ate Gas circuit clogged Uninstall and clean heat ex changer if necessary re place filter ele ment Condensate outlet iced over Send in cooler Error messages on the display The display alternates betwee...

Страница 12: ...s de pr vention des ac cidents sont respect es les donn es et conditions d utilisation licites sont respec t s les dispositifs de protection sont utilis s et les travaux d entretien prescrits effectu...

Страница 13: ...pompe p ristaltique Le raccord de 6 pour la sortie de condensat de la pompe est enficher d licatement et de la mani re appropri e au moyen du tuyau correspondant et du collier de serrage Les versions...

Страница 14: ...eur d humidit humidit restante en option C Analogique Sortie Temp rature 4 20 mA en op tion B Etat Refroidisseur sur temp rature ou sous temp rature D Masse de l appareil Rac cordement du blindage de...

Страница 15: ...C 20 C 35 6 F 68 F Placer la limite sup rieure d alarme dans la plage tEMP 1 C 7 C 1 8 F 12 6 F Placer la limite inf rieure d alarme dans la plage tEMP 1 C 3 C 1 8 F 5 4 F Si un capteur d humidit est...

Страница 16: ...s D bit de gaz diminu Voies de gaz bou ch es D monter l chan geur thermique et le nettoyer remplacer l l ment de filtre le cas ch ant Sortie de conden sat gel e Exp dier le refroi disseur Messages d e...

Страница 17: ...identes Se cumpla con los datos aportados y las condiciones de uso Se utilicen los dispositivos de seguridad y se lleven a cabo las tareas de mantenimiento exigidas Se tengan en cuenta las regulacione...

Страница 18: ...ica La conexi n 6 para la salida de condensados de la bomba de be establecerse por profesionales y con extremo cuidado utili zando una manguera y abrazaderas Las versiones con uniones roscadas DN 4 6...

Страница 19: ...los A Estado del sensor de hu medad humedad resi dual opcional C Salida de temperatura anal gica 4 20 mA op cional B Estado del refrigerador temperatura excesiva mente alta baja D Masa del aparato Con...

Страница 20: ...uperior en valores tEMP 1 C 7 C 1 8 F 12 6 F Establecer el l mite de alarma inferior en valores tEMP 1 C 3 C 1 8 F 5 4 F Si el sensor de humedad est instalado bot n de calibrado Si el sensor de humeda...

Страница 21: ...udal de gas reducido Conductos de gas atascados Desmontar y lim piar el intercam biador de calor en caso necesario reemplazar el ele mento de filtro Salida de conden sado cubierta de hielo Remitir ref...

Страница 22: ...0 Ratingen Deutschland 49 0 2102 4989 0 49 0 21 02 49 89 20 1 1 1 2 1 3 2 a b c d a b c 3 20 C 60 C 4 F bis 140 F 4 4 1 19 10 cm 4 2 IN 11 LD spez AK20 V 165 SS DN8 10 5 16 4 2 1 6 DN 4 6 1 6 1 4 CPsi...

Страница 23: ...1 kV 1 5 kV EN 175301 803 1 5mm2 AWG 16 8 10 mm 0 31 0 39 1 A100048 230 VAC 50 60 Hz 115 VAC 50 60Hz 10 A 8 10 mm 4 4 12 PHOENIX 250 VAC DC 1 A 1 2 2 12 PH NIX A C C 4 20 mA B D 4 20 mA 5 5 1 3 23 B...

Страница 24: ...H2O ____ ____ ____ E219 ____ ____ HSens ____ ____ 5s C F 2 C 20 C 35 6 F 68 F tEMP 1 C 7 C 1 8 F 12 6 F tEMP 1 C 3 C 1 8 F 5 4 F tEMP A Hi A Lo h2o hSEn E HiLo rSt ___ ___ ____ C toP E 6 CD www buehl...

Страница 25: ...EGK 2 19 7 1 20 Error 01 Error 02 1 2 1h2o 2h2o 1bw 2bw 25 B hler Technologies GmbH BX450010 10 2018...

Страница 26: ...hnologies com B hler Technologies GmbH Harkortstra e 29 40880 Ratingen Deutschland 49 0 21 02 49 89 0 49 0 21 02 49 89 20 1 1 1 2 1 3 2 a b c d a b c 3 20 C 60 C 4 F 140 F 4 4 1 19 26 B hler Technolog...

Страница 27: ...EGK 2 19 10 4 2 IN 11 LD AK 20 V 165 SS DN 8 10 5 16 4 2 1 jv 6 DN 4 6 1 6 1 4 CPsingle CPdouble 1 4 3 IEC 60947 1 IEC 60947 3 1 1 5 27 B hler Technologies GmbH BX450010 10 2018...

Страница 28: ...19 EN 175301 803 1 5 2 AWG 16 8 10 0 31 0 39 1 A100048 230 AC 50 60 115 AC 50 60 10 8 10 4 4 12 PH NIX 250 AC DC 1 A 1 2 2 12 PH NIX A C 4 20 A B D 4 20 5 5 1 3 28 B hler Technologies GmbH BX450010 10...

Страница 29: ...2O ____ ____ Top Settings ____ Exit E219 ____ ____ HSens ____ ____ Exit 5 C F 2 C 20 C 35 6 F 68 F tEMP 1 C 7 C 1 8 F 12 6 F tEMP 1 C 3 C 1 8 F 5 4 F High Low tEmP A Hi A Lo h2o hSEn E HiLo rSt ___ __...

Страница 30: ...EGK 2 19 7 1 20 01 02 1 2 1h2o 2h2o 1bw 2bw 30 B hler Technologies GmbH BX450010 10 2018...

Отзывы: