31
MA112780_38.6
[Original Instructions]
9/19/2011
Removing the Automatic Blade Dressing System
(Catalog Number 11-2696)
1. Turn OFF the IsoMet 5000.
2. Disconnect the power cable from the control housing.
3. Remove and keep the two (2) mounting screws.
4. Recap the bottom power socket.
Installing the Automatic Blade Dressing System
1. Turn OFF the IsoMet 5000.
2. Clean the top of the turret of all cutting materials and fluids.
3. Place the Automatic Blade Dressing unit with the mounting flange down and to the right.
4. Align the mounting holes.
5. Insert the screws and tighten.
6. Uncap the power socket receptacle.
7. Insert the power plug and secure it.
8. Attach a dressing stick using the thumbscrew.
9. Adjust the T-mount to the blade size being used.