21
MA112780_38.6
[Original Instructions]
9/19/2011
Operation
Cutting a Specimen and Serial Sectioning
To cut a specimen, select the appropriate BLADE SPEED, FEED RATE, and CUTTING LENGTH
parameters.
If using the Specimen Positioning System, the SAMPLE THICKNESS, SPECIMEN QUANTITY,
and BLADE THICKNESS must also be set.
All the necessary functions will automatically activate when the CUTTING CYCLE button is
pressed to the ON position. The CUTTING CYCLE command is accessible in the L1, L2, and L3
display screens.
WARNING
!
Equipment Damage. The IsoMet 5000 Linear Precision Saw is designed to only operate
with the cover closed. This prevents manual feeding of specimen materials or dressing
sticks into the rotating blade, which could result in personal injury and/or blade damage.
Efforts to defeat the safety interlock could result in personal injury and void warranty.
Single Cut
without
the Specimen Positioning System
1. Adjust the BLADE SPEED, FEED RATE, CUTTING LENGTH, SAMPLE THICKNESS, and
BLADE THICKNESS.
Press the SCROLL button until the desired parameter is highlighted and adjust the
parameter values using the INCREASE or DECREASE buttons.
2. Adjust SPECIMEN QUANTITY to 1.
3. Position the specimen using any of the required chucks.
4. Use the SCROLL button to scroll to the L1 Screen.
5. Press the PUMP MOTOR button to the ON position and position the coolant flow as desired
6. Press the CUTTING CYCLE button to the ON position. The blade will begin to rotate at the
selected speed, the coolant will begin to flow, and the automatic feed will begin to advance.
As the blade advances the numerical value on the display screen will start to decrease.
7. When the DISTANCE REMAINING field reaches .00 the blade turret will retract and turn
OFF.