![Buehler 11-2675 Скачать руководство пользователя страница 34](http://html.mh-extra.com/html/buehler/11-2675/11-2675_manual_2826462034.webp)
28
MA112680_39.3
[Original Instructions]
4/2014
Troubleshooting Chart
Problem
Possible Cause
Correction
The IsoMet does not power on
The unit is not plugged in
Check the power cord
connection
The unit is not powered on
Check the rear power switch
The pump has lost pressure
Fluid level may be too low
Fill the fluid to the correct level
Tank screen may need cleaning
Remove screen and clean
Suction hose clogged
Remove and check suction and
flexible hose for blockages and
clear blockages
Flexible hoses clogged or
blocked
The blade keeps pinching
Wrong blade for the material
Replace with correct blade
Improper specimen fixture
(primary cause)
Use correct specimen fixture
Blade may need to be dressed
Dress blade
The FEED RATE may be too
slow
Increase FEED RATE
BLADE SPEED may be too slow
Increase BLADE SPEED
The blade keeps stopping
The CUTTING DISTANCE may
have changed
Reset the CUTTING DISTANCE
The DISTANCE REMAINING
may be .00 or less
Increase the DISTANCE
REMAINING
The ARM LIMIT switch may be
activated
Deactivate the ARM LIMIT switch
The bed has rust on it
Improper coolant mixture
Check the coolant for proper
mixture
The hood has been closed too
long with moisture build up
Open hood and keep hood open
when not in use
The Specimen Positioning
System does not move
Cable is disconnected
Check the cable connection
The Blade Dressing does not
work
Cable is disconnected
Check the cable connection
Dressing stick in the wrong
position
Check the dressing stick position
and clean it
The Rotating Chuck does not
work
Cable is disconnected
Check the cable connection
Rotating chuck is not powered on
Make certain the Rotating Chuck
is turned on
The Blade Turret completed one
cycle but won’t advance for a
second cut
The IsoMet is operating as if the
Precision Table is still attached
Turn the unit off then back on.
Coolant leaks from the IsoMet
Table or unit is not level
Make certain the table is level
Coolant level too high
Check the volume of cutting fluid