![Buehler 11-2675 Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/buehler/11-2675/11-2675_manual_2826462015.webp)
9
MA112680_39.3
[Original Instructions]
4/2014
Scroll
The SCROLL button scrolls through the different parameter fields highlighting
each one as it is selected. The parameter fields will change depending on how many
times the SCROLL button is pressed.
There is a total of four (4) parameters that can be displayed: BLADE SPEED, FEED
RATE, DISTANCE REMAINING, CUTTING DISTANCE.
Zero
The ZERO button is used to indicate the SOFT HOME position by positioning
the blade and the Micrometer Positioning System (if attached) at a starting point just
before the specimen.
Once a position is determined, press the ZERO button and the DISTANCE
REMAINING value will change to .00, indicating the SOFT HOME position.
Increase
The Increase button will incrementally increase (raise) a parameter’s
value. The maximum values for the IsoMet 4000 are:
•
BLADE SPEED = 5000 rpm
•
FEED RATE = .75 in/min [19 mm]
•
CUTTING DISTANCE = 8.00 inches [203 mm]
Decrease
The Decrease button will incrementally decrease (lower) a parameter’s
value. The minimum values for the IsoMet 4000 are:
•
BLADE SPEED = 200 rpm
•
FEED RATE = .05 in/min [1.27 mm]
•
CUTTING DISTANCE = .01 inch [.26 mm]
EMERGENCY STOP
Is the big red knob on the right side of the front panel. When pressed all electrical
power is disconnected from the blade and all moving parts, disabling any further
operations.
A warning message will appear when the Emergency Stop button is pushed.
To return power, turn the Emergency Stop knob clockwise.
Press the CUTTING CYCLE button to continue operation.