background image

DCCT Pinhole Detector

 

5

 

 

 

 

Unpacking 

 

Remove the DCCT and its accessories from the packaging and check for 
any damage. Should any item be damaged, notify the supplier and carrier 
immediately. Keep all packing material for inspection. Do not use the 
equipment. 

 

Specifications and symbols 

 

DCCT unit Type

 

Output test voltage (DC neg.)

 

 

DCCT 1 

 7.5 kV 

Adjustable 1kV to 7.5kV 

 

DCCT 2 

 15 kV 

Adjustable 2kV to 15kV 

 

DCCT 4 

 30 kV 

Adjustable 4kV to 30kV 

 

DCCT 5 

 40 kV 

Adjustable 5kV to 40kV 

 

Sensitivity: 

Variable (factory set for BS EN 62230). 

Output current: 

Maximum continuous output current: <1mA. 

Operating time: 

Continuous 

Weight: 

17 Kg 

Dimensions: 

500mm (W), 300mm (H), 330mm (D) 

Meter accuracy: 

<2%. 

Mains supply voltage:  110/120V or 220/240V, 50/60Hz. 
Standard: 

Conforms to BS EN 62230 - Electric cables, 

Spark-test method. 

 

The following symbols are displayed on the DCCT equipment for your safety 
and protection. Failure to comply with the instructions contained in this 
manual could result in damage to the equipment or personal injury. 

 

 

 

Caution, risk of danger 

Caution, risk of shock 

Chassis terminal

 

 

NOTE: 

It is important that you read this manual fully to familiarise yourself 

with the DCCT unit. 

Содержание DCCT

Страница 1: ...MANUFACTURERS OF SPECIALIST TEST EQUIPMENT DCCT Pinhole Detector Operating Instructions English Version Please read carefully before use...

Страница 2: ......

Страница 3: ...Voltage Connections 6 Alarm Circuit 6 Slow Speed 6 Fast Speed 6 Changing pre set speed 7 Remote operation 8 Operating instructions 9 Maintenance 9 Safety precautions 10 Electromagnetic compatibility...

Страница 4: ...can be adjusted by inserting a small screwdriver into the aperture on the left hand side of the front panel marked ALARM DELAY to an give audible visual warning from 0 5 to 4 seconds duration The ele...

Страница 5: ...vity Variable factory set for BS EN 62230 Output current Maximum continuous output current 1mA Operating time Continuous Weight 17 Kg Dimensions 500mm W 300mm H 330mm D Meter accuracy 2 Mains supply v...

Страница 6: ...minal at the rear of the DCCT unit and the conductive substrate of the item to be tested This will ensure a complete earth connection and completes the circuit for detection of faults Note Failure to...

Страница 7: ...arm speed mode switch for changing the mode from SLOW SPEED to FAST SPEED is at the back of the circuit board Fig 2 Replace and secure the unit in its case and re connect the leads as necessary The AL...

Страница 8: ...rocedure if remote operation is no longer required A set of heavy duty change over contacts are provided for operating exterior contactors solenoids etc Connection is made to terminals 1 2 and 3 of CO...

Страница 9: ...e operating and also that the test voltage drops to zero Ensure that the test area is dry and clean Sweep test area with an electrode at a steady speed and observe for any drop of indicated test volta...

Страница 10: ...he will receive an electric shock Whilst not in itself hazardous the involuntary reaction to an unexpected electrical shock may cause a related accident i e stepping backwards into an adjacent piece...

Страница 11: ...tic appliances and household electronic equipment Aeronautical and marine radio apparatus Educational electronic equipment Telecommunications networks apparatus Radio television broadcast transmitters...

Страница 12: ...than expected Remember that electromagnetic emissions travel in all directions and can even be bent or deflected Neighbouring works or sites should also be taken into account there may be equipment in...

Страница 13: ...r leads are used they could transmit similar to aerials and cause interference Wherever you are intending to use equipment of this type on your site or on a customers always obtain clearance from the...

Страница 14: ...Instruction Manual 14 Output graphs The following are the output graphs for the four models of DCCT available Please review the graphs associated with your DCCT model...

Страница 15: ...DCCT Pinhole Detector 15...

Страница 16: ...Instruction Manual 16...

Страница 17: ...DCCT Pinhole Detector 17...

Страница 18: ...ity We Buckleys UVRAL Ltd As manufacturer of the apparatus listed declare that the product DCCT DC Holiday Detector Is manufactured in conformity with the following directives 2014 30 EU 2014 35 EU Au...

Страница 19: ...Manufacturer Buckleys UVRAL Ltd Address Buckleys House Unit G Concept Court Shearway Business Park Shearway Road Folkestone Kent CT19 4RG UK Tel 44 0 1303 278888 Fax 44 0 1303 274331 Website www buckl...

Страница 20: ...s relating to your product Be the first to hear about our new products We strive to improve the quality of our products and service Registering your product helps us monitor overall quality of our pro...

Отзывы: