6 Operation
34
K-439
Operation Manual, Version B
Starting an automatic digestion by loading a method
• Select Main Menu > Methods.
• Highlight the method that should be loaded by using the up or down buttons.
• Press >>.
• Now press Load. The “Automatic” screen appears.
• Select the row(s) to be active, by pressing Row and then the ± buttons. Then press OK to confirm.
• To define an optional delay time for the start of the digestion, press Timer and enter the corre-
sponding time (in hh:mm) using the ± buttons. Then press OK to confirm. The preheat function is
automatically set to "Preheat ..continue" when the timer is activated.
• To change the Preheat function, press Preheat and select "..wait" or "..continue" using the ±
buttons. Then press OK to confirm.
• Press the START button to start the preheating. As soon as the temperature entered for preheating
is reached, the system is ready to start the digestion. The display starts blinking and the buzzer
beeps three times to indicate the ready status.
• When the preheat function "Preheat ..wait" is selected, press the START button to start the diges-
tion process.
When the preheat function "Preheat ..continue" is selected, the digestion process starts directly.