6 Operation
32
K-439
Operation Manual, Version B
Note
If the rack(s) is/are cooled down inside the heating chamber, the time for the cool down step in the
corresponding method must be set to
≥
80 min, so that no harmful vapors can escape. If the racks
are cooled outside the heating chamber, the cool down step in the corresponding method must be
set to
≥
30 min.
• Press Save.
• Now enter a name for the new method by highlighting individual letters on the virtual keyboard with
the ± buttons and press Enter to confirm a letter.
• Repeat this procedure until the name is completed.
• Press Save.
The new method now appears in the method list and can be loaded to start a digestion.