5 | Installation
NIR-Online GmbH
24/36
Operation Manual Process Analyser X-One
5.7
Connecting the sensor
Tools required:
Torque wrench, size 7 mm AF
Torque wrench, size Torx T20
NOTE
Make sure that the power is not switched on when connecting the sensor.
NOTE
Loss of performance due to use of unsuitable device cables
Max. cable length between installation box and sensor 10 m.
u
Slide the cable guard over the connec-
tor.
u
Plug the connector into the sensor.
u
Secure the connector.
Tightening torque: 2.5 Nm ± 0.5
u
Fix the cable guard to the sensor.