8 | Preparations for a determination
Büchi Labortechnik AG
74/106
Operation Manual K-365 Kjel Line
8.11
Preparing the level detection sensor (option)
Precondition:
R
The installations for the level sensor are
completed. See Chapter 5.18 "Installing the
level detection sensor (option)", page 37
u
Attach the magnetic level sensor to the metal
plate.
10 mm
u
Adjust the sensor near the sample receiving
vessel.
8.12
Selecting the operation mode
Select one of the following operation modes:
Operation mode
Explanation
[Single Determination]
For carrying out one determination. Editing settings see
Chapter 8.3 "Editing the single determination", page 52
[Series]
For planning and carrying out a series of determinations.
Editing settings see Chapter 8.4 "Editing a series", page 55
Specifications on the display:
u
Navigate to the specified menu.
u
Select the specified action.
...
Operation Mode
u
Tap the specified function on the function bar.
EDIT
u
Select the operation mode according to your
requirements.
...
...
...
...
u
Tap the specified function on the function bar.
SAVE
8.13
Dosing manually
8.13.1 Dosing H₂O manually
NOTE
The quantity of liquid per keystroke can be changed in the
[Settings]
submenu.