Büchi Labortechnik AG
Preparations for a determination | 8
Operation Manual K-365 Kjel Line
63/106
Specifications on the display:
u
Select the specified action.
...
H₂SO₄ for Digest.
u
Tap the specified function on the function bar.
EDIT
u
Enter the quantity of the sulfuric acid.
...
u
Tap the specified function on the function bar.
ð
The quantity for the sulfuric acid is saved.
SAVE
8.6.7
Changing the NaOH volume for a method (MultiKjel only)
NOTE
The
[Reaction Detection]
action is set to Off. See Chapter 8.6.4 "Changing the
reaction detection for a method (option for MultiKjel only)", page 61
Enter manually the quantity of NaOH for alkalizing the sample.
For the correct quantity check related
"Application note "
or use the KjelOptimizer
App.
Specifications on the display:
u
Navigate to the specified submenu.
Methods
u
Select the method that you wish to edit.
...
u
Select the specified action.
...
NaOH Volume
u
Tap the specified function on the function bar.
EDIT
u
Enter the quantity for the dosing.
...
u
Tap the specified function on the function bar.
ð
The quantity for the dosing is saved.
SAVE
8.6.8
Changing the acid volume for a method (BasicKjel, MultiKjel only)
Enter manually the quantity of reagent to be dosed.
Specifications on the display:
u
Navigate to the specified submenu.
Methods
u
Select the method that you wish to edit.
...