USER SETTINGS OPTION | PAGE 36 OF 62
4.1.2
PARAMETERS OF SECTIONS IN SEQUENCE
A sequence consists of a few currents / programs that are called sections. Parameters of sections must be set
when creating a sequence.
Each program includes basic current parameters such as frequency, pulse length, modulation, etc. For more
information, please refer to the chapter
Save Therapy
. Set all data in the manual setting screen and save them
as a user-designed program (diagnosis). Insert the program/diagnoses in the sequence. Set the length of time of
the section when inserting the program/diagnoses in the sequence (except laser, where the time of the section
depends on the currently connected laser probe). Obviously, the factory-preset programs can also be inserted in
the sequences. In the section only the polarity can be set (for electrotherapy sequence). The other parameters
must be specified and saved in the inserted program.
Example: you want to create a sequence of diadynamic DF current (without base, positive polarity, CC mode,
time of stimulation: 1 minute) and CP-ISO current (base 10%, reversal of polarity in the middle of the set time, CC
mode, 10 minutes). Press
man
to select the manual mode, set diadynamic currents, DF type, without base,
positive polarity, CC mode. Save this setting as (for example) program E-8001. Then set the parameters of the
CP-ISO current: base 10%, positive, reversal, CC mode, and save it as (for example) program E-8002. Select
therapy and press
new sequence
then press
new
, set the program number
8001
, set the time of section
1:00,
positive polarity, and press
enter
. Then add the second section in the same way –
new
, program number
8002
,
time of section
10:00
, positive polarity with reversal, and press
enter
. Then press
enter
again to return to the
manual settings screen, press
save/copy
and save the sequence (for example, as number 9501). The cv/cc
mode is set globally for the whole sequence before starting it.
4.1.3
SAVING NEW SEQUENCE
Sequence created according to the chapter
Creating New Sequences
can be saved as follows:
Either directly from the screen for setting of new sections.
Or from the main manual screen in the same way as the new diagnosis.
User-made sequences are saved under numbers
9500
-
9999
. They can be found in the list of programs, in the
list of diagnoses or in the list of sequences.