BTL BTL-09 Скачать руководство пользователя страница 19

BTL - 09 

 

user manual & user's guide 

page 19 of 42 

Applicators 

Aplicators for BTL-09 are manufactured from top-quality materials that provide a superior surface finish for easy and 
non-demanding maintenance. The patient does not have to disrobe for therapy. In addition, neither bandages nor 
plaster weakens the magnetic field. 
Applicators are manufactured in three general types: 
 

1. 

 

Universal (mats, round, square, etc.) 

2. 

 

Solenoids (cylinders of various width - 50 and 30 cm in diameter) 

3. 

 

Desk (solid, inflexible applicators up to 3 cm in width) 

Magnetic field 

BTL-09 generates rectangular pulses. For easy use, the most frequently used therapies are pre-programmed in BTL-
09. Experience has proved that a slow frequency of 3.5 to 7.0 Hz is best for inflammatory states and acute painful 
states, with the frequency of 14 to 28 Hz for chronic states. 14 Hz is most suitable for application to the head, the 
spine, or the whole body for neurological ventricular diseases. 
  
If you use the pulse magnetic field for acute or chronic pain, there may be a temporary worsening of the pain. This is 
often caused by over supplying the tissue with blood in combination with an activation of the inflammation. This is also 
an indicator of the patient’s reaction to the pulse magnetic field, which is a positive sign. The application length is 
usually around 30 minutes a day, and should not be interrupted even if a temporary worsening of pain occurs. In such 
a case, you can lower the pulse magnetic field’s intensity by 50% for the following 3-4 days, and then return to the 
original level of intensity.

 

When the first signs of pain lessening appear (the back or joints), continue to apply the pulse magnetic field for 
another 3-4 days, and then end the treatment. Usual therapy period is 2 to 3 weeks. For neurological diseases, a long 
term (even for several months) interrupted application should be expected. 

Contraindications 

 

absolute contraindication during pregnancy (either patient or therapist) 

 

acute fevers, general tiredness, active TBC 

 

all bleeding states 

 fungi 

diseases 

 

patients with a pacemaker 

Combination with other physical methods 

Pulse magnetotherapy can be combined with all other physical methods (ultrasound, laser, iontophoresis, diadynamic 
current). It is also very suitable to combine with an intensive rehabilitation, water treatment, acupuncture, 
physiotherapy, etc. 
 
The combined use of antibiotics and the pulse magnetic field has proved very efficient, for it leads to amplification of 
the antibiotics that had previously been ineffective. This effect seems to be especially strong with bone inflammations. 
The application of the pulse magnetic field and corticoids is also very efficient. For example an application of Kenalog 
with mesocain into the joint or soft tissues displays a healing effect for 3-4 weeks, but when combined with a pulse 
magnetic field the effect is extended for a period of a number of months and even up to one year. When using 
Diprophos rather than Kenalog, the resultant beneficial effect is much longer. 

Methodology 

BTL-09 offers a wide range of possibilities for the choice of an optimal therapeutic procedure, whether it be the wide 
selection of pre-programmed settings or the ability to directly set the pulse magnetic field frequency. The flexibility of 
BTL-09 is very much appreciated by pulse magnetotherapy specialists. 
 
For simple and safe everyday operation, programs have been pre-selected -- based on long-term experience -- that 
have the best effects in most cases. Each generated pulse magnetic field has its specific bio-trophic parameters that 
evoke a specific reaction in a living organism. Any change in these parameters can result in an unexpected negative 
effect. This includes primarily the affecting of skeletal conditions. For example, instead of an expected muscle creation 
in a badly self-healing fracture, inappropriate methodology can cause the dissolving of calcium ions into soft tissue. 
This can sometimes form large deposits of soft tissues in the area of joints, and worsen the overall condition. On the 
other hand a clinically monitored 2.5-year experiment has proved that there are a few people who have absolutely no 
reaction to any magnetic field, even with a high intensity. 
 
Experience shows that a temporary worsening during the 3

rd

-5

th

 day after beginning treatment does not mean that 

therapy should be interrupted. On the contrary, it is a good sign that the patient is positively reacting to the pulse 
magnetic field. This is because the irradiated tissue is, due to vasodilatation, oversupplied with blood and as a result, 
subjective conditions become temporarily worse. It is also necessary to explain to the patient that when therapy is 

Содержание BTL-09

Страница 1: ...v3 20z8JD10 03 2005EN user manual user s guide...

Страница 2: ...ns 7 Warranty 8 II APPLICATORS 9 Instructions for use 9 Technical data 9 Field outside application area 9 Used material 9 Big solenoid 1 10 Small solenoid 2 11 Ring 3 12 Disc 4 13 Double cover 5 14 Tr...

Страница 3: ...tput 15 Control lights to signal output failure OUTWARD APPEARANCE BTL 09 is built in a lacquered metal box Signals displays controls and connectors for applicators or adapter plug are located on the...

Страница 4: ...ven later in this manual Code Description 1 Big solenoid 50 cm diam 2 Small solenoid 30 cm diam 3 Ring 30 cm diam 4 Disc 24x24 cm 5 Double cover 2x24x17 cm 6 Triple cover 59x28 cm 7 Adapter plug Plug...

Страница 5: ...ime has elapsed INTENSITY SETTING Intensity of magnetic field is shown on Intensity display 12 Intensity displayed is the maximum intensity at applicator s surface Intensity around individual applicat...

Страница 6: ...push button there may not be any therapy going on and the control light Channel 6 of relevant channel must be on To switch to another channel press the push button 5 SHORT INSTRUCTIONS FOR USE 1 Conn...

Страница 7: ...defect remains only at one channel you can use the second channel as an emergency No applicator may be connected to the faulty outlet TECHNICAL SPECIFICATIONS Number of independent outputs 2 Magnetic...

Страница 8: ...or fax the fully completed warranty registration form within 30 days of installation All costs of shipment are the responsibility of the purchaser Damage to any part such as by accident or misuse or i...

Страница 9: ...use Have the applicator checked every 6 months by a qualified service person TECHNICAL DATA Field outside application area The cylinder coils i e both solenoids and ring do not have their correspondi...

Страница 10: ...of hard polystyrene that forms additional insulation Internal part of applicator is upholstered Movable outlet is provided by rubber outlet Fig 1 shows levels of magnetic induction valid for all inte...

Страница 11: ...f hard polystyrene that forms additional insulation Internal part of applicator is upholstered Movable outlet is provided by rubber outlet Fig 1 shows levels of magnetic induction valid for all intens...

Страница 12: ...f supporting coil with double insulation Fig 1 shows setting of levels of magnetic induction valid for all intensity values The numbers relate to maximum actuation of applicator The higher values rela...

Страница 13: ...flat coil built in to both side upholstered pillow Fig 1 shows setting of levels of magnetic induction valid for all intensity values The numbers relate to maximum actuation of applicator The higher...

Страница 14: ...magnetic induction levels valid for all intensity values The numeric values relate to the maximum actuation of applicator The higher values relate to use without adapter plug the lower values relate...

Страница 15: ...BTL 09 user manual user s guide page 15 of 42 max intensity if adapter plug is used max intensity Fig 1c Double cover Fig 2 Double cover...

Страница 16: ...makes together with internal parts relatively rigid structure which is still partially flexible The coils are poled alternatively Fig 1 shows setting of magnetic induction levels valid for all intens...

Страница 17: ...immunity system or changes in fagocytosis etc Electromagnetic waving means the spreading of a variable magnetic field through space and it carries energy The irradiated tissue undergoes a number of ph...

Страница 18: ...netic field improves information transfer of nerve stimulation muscle ability to contract blood circulation erythropoiesis stimulation improves tissue pH improves the immunity system through hypophysi...

Страница 19: ...very suitable to combine with an intensive rehabilitation water treatment acupuncture physiotherapy etc The combined use of antibiotics and the pulse magnetic field has proved very efficient for it l...

Страница 20: ...not to be shortened all complications will not necessarily disappear immediately Instead symptoms may fade away gradually after therapy has been completed After the healing process has been triggered...

Страница 21: ...aena 26 Morbus Perthes 27 Myalgia 28 Myopathia 29 Neurodermatitis 30 Osteochondrosis 31 Osteomyelitis acuta chronica 32 Osteonecrosis aseptica 33 Osteoplastica 34 Osteoporosis 35 Osteosynthesis 36 Ost...

Страница 22: ...here is a clinical finding but no morphological change has yet occurred apply magnetotherapy to the entire body in combination with non steroid anti rheumatoid medicaments such as Voltaren or Diclofen...

Страница 23: ...15 Duration 30 45 minutes a day Application max 9mT for 45 minutes or max 11mT for 40 minutes Apply according to a location mostly big solenoid max 13mT for 20 minutes 11 Degeneratio spinalis Program...

Страница 24: ...actura fractures in casts or with an osteotomy Program 15 or 19 Duration 30 max 40 minutes per day Therapy can be repeated Application max 11mT program 15 max 7mT program 19 for UL and LL small soleno...

Страница 25: ...ion 9 12 mT for applicators other than solenoids 4 6 mT for solenoids Place the applicator mainly in the spinal area using a solenoid or double cover for feeble patients 29 Neurodermatitis Program 14...

Страница 26: ...ay For example 3 times a day max for 90 minutes per session Application 3 5mT Comment It is always necessary to remove sequesters and evacuate hollows with a pus content Plaster or metal implants have...

Страница 27: ...lly improves mood and imparts a sense of well being However expect a temporary worsening after 2 or 3 applications Do not change the program or cease treatment but intensity can be lowered 50 and retu...

Страница 28: ...am 24 max 18mT for 30 minutes 53 Sy vertebralis painful vertebral syndrome with muscle spasms with nerve root stimulation Program 8 or 30 Duration 30 minutes Application triple cover program 8 max 14m...

Страница 29: ...utes for ring Comment Very effective in early stages to decrease inflammation of the mucous membranes 60 Periostitis Program 1 Duration 5 8 therapy sessions for 30 minutes Application 6mT Use of Appli...

Страница 30: ...lses 20ms with 20ms break frequency 25Hz Program 2 20x pulse 10ms with 10ms break break 100ms Program 3 10x pulse 20ms with 20ms break 150ms break Program 4 60ms pulse with 60ms break 5x pulse 10ms wi...

Страница 31: ...pulse with 100ms break Program 8 30ms pulse with 60ms break frequency 11 1Hz Program 9 3x 20ms pulse with 20ms break 60ms break 10x 5ms pulse with 5ms break 120ms break Program 10 10ms pulse with 20ms...

Страница 32: ...equency 3 5Hz 4x 71ms pulse with 71ms break frequency 7Hz 8x 35ms pulse with 36ms break frequency 14Hz 13x 5ms pulse with 30ms break frequency 28 5Hz Program 14 10ms pulse with 10 23ms break with repe...

Страница 33: ...ith 212ms break frequency 4 6Hz Program 18 3x 35ms pulse with 115 150 break 15x 5ms pulse with 145ms break frequency 6 6Hz Program 19 15ms pulse with 40ms break frequency 18 2Hz Program 20 3x 65ms pul...

Страница 34: ...4x 35ms pulse with 35ms break 35ms pulse with 80ms break 4x 18ms pulse with 18ms break 18ms pulse with 500ms break Program 24 10x 9ms pulse with 9ms break 100ms break 40ms pulse with 80ms break 2x 40...

Страница 35: ...equency 6 6Hz Program 28 15ms pulse with 28ms break frequency 23 25Hz Program 29 5x 5ms pulse with 5ms break frequency 100Hz 80ms Program 30 20ms pulse with 200ms break frequency 4 5Hz Program 31 10x...

Страница 36: ...frequency 25Hz 406ms 3x 16ms pulse with 510ms break frequency 1 9Hz Program 35 3x 20ms pulse with 20ms break frequency 25Hz 183ms break 4x 5ms pulse with 298ms break frequency 3 3Hz Program 36 5x 6ms...

Страница 37: ...uency 23 3Hz 189x 7ms pulse with 30ms break frequency 27Hz 220x 15ms pulse with 15ms break frequency 33 3Hz Program 41 10x 5ms pulse with 5ms break 212ms break 10x 10ms pulse with 10ms break 13x 5ms p...

Страница 38: ...17ms pulse with 44ms break frequency 16 4Hz Program 45 5x 6ms pulse with 4ms break 20ms break Program 46 7x 30ms pulse with 20ms break 217ms break 12x 5ms pulse with 212ms break Program 47 5x 40ms pul...

Страница 39: ...476ms break 3x 5ms pulse with 471ms break frequency 2 1Hz Program 52 10x 5ms pulse with 10ms break frequency 66 6Hz 100ms 60ms pulse with 80ms break 60ms pulse with 100ms break 60ms pulse with 80ms b...

Страница 40: ...with 295ms break frequency 3 3Hz Program 56 10ms pulse with 155ms break frequency 6 1Hz Program 57 5x 20ms pulse with 100ms break frequency 8 3Hz 1065ms break Program 58 5x 7ms pulse with 4ms break 9...

Страница 41: ...th 15ms break frequency 50Hz Program 63 100x 5ms pulse with 35ms break frequency 25Hz 100x 5ms pulse with 5ms break frequency 100Hz Program 64 5ms pulse with 20 45 ms break frequency 20 40Hz 120s peri...

Страница 42: ...any means incl electronic mechanic photographic or other records without previous approval from BTL Industries Limited BTL Industries Limited operates a policy of continuous development Therefore it r...

Отзывы: