BTL - 09
user manual & user's guide
page 14 of 42
DOUBLE COVER (5)
Dimensions:
240x170x20 mm
Weight: 3.7
kg
Inductance: 33
mH
Resistance: 4.6
Ω
Max. voltage:
80 V
Max. current:
6 A
Max. power:
30 VA
Max. temperature:
45 C
Insulation: class
II
Environment: standard
The applicator consists of two mats (covers). The connection is provided by belt of 170mm length. Pillows consist
of elliptical self-supporting coils insulated and upholstered from both sides.
Fig. 1 shows magnetic induction levels valid for all intensity values. The numeric values relate to the maximum
actuation of applicator. The higher values relate to use without adapter plug, the lower values relate to use
with adapter plug. Fig. 1a shows double cover with its two pillows spread out, fig. 1b shows the position of one cover at
50 mm distance from the other, and fig. 1c shows double cover with pillows at distance given by the length
of the connecting belt.
Electrical current induced in tissues is in direct relation to the speed of change of magnetic induction shown at fig. 2.
m
ax
. i
nt
ens
ity
if
a
da
pt
er
pl
ug
is
us
ed
Fig. 1a and Fig. 1b: Double cover