6.3 Outlet side 

The pump have integrated exhaust filters which, even at a high gas throughput, trap 

the oil mist and guarantee exhaust gas free of oil mist. If the exhaust filters are clogged, 
pressure relief valve opens and the filters are bypassed. As a result, the proportion of oil in 
the exhaust gas as well as the pump’s oil consumption rises. Installing new exhaust filters 
will correct this problem.   

This situation must be resolved by changing the exhaust filters. It is under utilizer’s 

responsibility to assure maintenance on materiel to avoid any  oil mist  of the limits 
authorized by regulations. 

Check in the individual case whether an exhaust line is necessary and/or prescribed. 

Volatile substances can pass through the filter. Depending on the processed gas, we 
recommend connecting an exhaust line; this is always necessary when the exhaust gases are 
dangerous.   

Observe the safety precautions that apply to your application and process gases. The 

pump’s exhaust port also has an internal thread. A hose can be connected via a suitable 
screw-in nipple 

The cross-section of the exhaust line should be at least the same as the pump’s exhaust 

port. If the exhaust line is too narrow, overpressure may occur in the pump. 

Install the exhaust line with a downward slope to prevent condensate from flowing 

back into the pump. If this is not possible, we strongly recommend to install a condensate 
trap . if several pumps are connected to one exhaust line, ensure an adequate cross-section 
and a non-return valve at the exhaust of each pump. 

Never  operate the pump with a blocked or restricted exhaust line. Before start-up 

ensure, that any blinds or similar shut-off devices in the exhaust line on the pressure side 
are opened and that the exhaust line is not obstructed. Such restrictions reduce the 
pumping speed, increase the temperature, and could overload the motor or cause a 
dangerous overpressure in the pump. Excessive pressure in the pump could damage the 
seals, blow out the sight glass, or destroy the pump housing. In addition to this explosion 
hazard, excessive backpressure can result in hazardous process gases leaking out of the 
pump. If you are purging the oil casing with inert gas, limit the inert-gas flow. 

When pumping dangerous gases, exhaust line must be tight   

The maximum exhaust pressure  must not exceed 1.15 bar (absolute). Also reliably 

prevent the occurrence of any blockage in the exhaust line. Exhaust filter, accessories and 
the tubing must be rated according to the maximum throughput. The maximum 
throughput is equivalent to the pumping speed of the pumps. 

 

Содержание SRV300

Страница 1: ...Instruction Manual VERSION 1 0 Single stage oil sealed rotary vane vacuum pump SRV300...

Страница 2: ...ping of Condensable Gases and Vapors 9 6 7 3 Cyclic process 10 6 8 Stop 11 6 9 Ultimate pressure 11 7 Maintenance 12 7 1 Maintenance schedule 12 7 2 Checking the oil 13 7 2 1 Oil level 13 7 2 2 Oil co...

Страница 3: ...or ordinances Warning Follow the safety instructions below and take appropriate precautions Otherwise it may cause personal injury and equipment damage All power sources connected to the pump must be...

Страница 4: ...ith explosion danger to avoid injury or fire Do not place flammable materials around the motor and the pump body to prevent fire Do not place any obstructions that obstruct ventilation around the moto...

Страница 5: ...n fine filtering occurs in the exhaust filter elements The proportion of oil in the exhaust gas is thus reduced below the visibility threshold over 99 entrapment rate Oil trapped in the exhaust filter...

Страница 6: ...pump Pumps are not designed for pumping of aggressive corrosive flammable or explosive gases These pumps are not designed for working in flammable or explosive environment The pump is not suitable fo...

Страница 7: ...5 Installation diagram 7...

Страница 8: ...e line is too narrow it reduces the pumping speed We recommend applying either LOCTITE or TEFLON tape to the screwed unions so that they are vacuum tight especially if gases are dangerous Pump should...

Страница 9: ...overpressure may occur in the pump Install the exhaust line with a downward slope to prevent condensate from flowing back into the pump If this is not possible we strongly recommend to install a conde...

Страница 10: ...ician in accordance with the applicable safety regulations Wire the motor for the correct supply voltage via connections in the junction box See the wiring diagram on the motor name plate For proper c...

Страница 11: ...the intake to check vacuum Exposure of a part of the body to the vacuum result in a rush of blood in the exposed part 6 7 Operation The pump gases and vapors provided that the gas ballast valve is ins...

Страница 12: ...le vapors until it has warmed to operating temperature pumping process gas with a cold pump results in vapors condensing in the oil For processes with a high proportion of condensable vapors slowly op...

Страница 13: ...ust be cooled down to the ambient temperature and must only be switched on again manually after having eliminated the cause In order to prevent the pump from running up unexpectedly after a main power...

Страница 14: ...nt conditions and or aggressive media may necessitate more frequent maintenance Maintenance job Frequency Section Check the oil level daily 7 2 1 Check the oil condition Depends on process 7 2 2 1st o...

Страница 15: ...operating temperature Depending on the process involved dangerous substances may escape from the pump and oil Take the appropriate precautions 7 3 Oil Change Always change the oil when the pump is sw...

Страница 16: ...that the filter element spring is behind the W shaped filter element bracket 6 Slowly insert the two positioning pins on the exhaust port cover into the corresponding holes in the W shaped filter ele...

Страница 17: ...mm away from the bottom side of the intake port Depending on the process involved dangerous substances may escape from the pump and oil Take the appropriate precautions 7 7 Radiator cleaning Please ke...

Страница 18: ...he pump The gas ballast must be closed and if the pump is to be shelved for a longer period of time it should be sealed in a plastic bag together with a desiccant If the pump has been shelved for more...

Страница 19: ...red without our company s consent 5 Failures and defects caused by abnormal working environment strong electromagnetic radioactive environment high temperature high humidity flammable gas environment...

Страница 20: ...ernal leak 1 Piping fittings loose Repair the pump Retighten Anti suck back valve is malfunctioning Repair the valve Exhaust valve is malfunctioning Repair the valve Float valve does not close Repair...

Страница 21: ...n intake line or in vacuum vessel Oil comes from the vacuum system Check the vacuum system Anti suck back valve is obstructed Clean or repair the valve Sealing surfaces or anti suck back valve are dam...

Страница 22: ...l elapsed time about 24hr continuous operation Intermittent operation Whether exposure to pollutants Yes No Name and Chemical formula pollutants Nature and character of pollutants Cleaning solvents an...

Страница 23: ...uipment Co Ltd Add No 55 Juchao Road Jiangkou subdistrict Fenghua District Ningbo City Zhejiang Province China Post code 315500 Tel 0574 88569588 59516555 Fax 0574 88569596 Email bsvac cnbaosi com Htt...