  Installation and Operation 

6.1 Installation 

It is essential to observe the following instructions step by step to ensure a safe start-up. 
Start-up may only be conducted by trained specialists. The pump can be set up on any flat, 
horizontal surface. Under the four feet, there are metric threaded holes for securing the 
pump. The oil level cannot be read properly if the pump is tilted. Pump risks running dry. 
The ambient temperature should be between 12

°

C (55

°

F) and 40

°

C (104

°

F) depending 

on oil type. To ensure adequate cooling of the pump, leave enough space at the air intake 
and exhaust points, so as for access and maintenance. Make sure to keep the air intake of 
motor clear. 

 

6.2 Inlet side 

The pump has an internally-threaded intake flange. Using suitable connecting elements 

the pump can be connected to the vacuum system. The cross-section of the intake line 
should be at least the same as the intake port. If the intake line is too narrow, it reduces the 
pumping speed. We recommend applying  either LOCTITE or TEFLON tape to the screwed 
unions so that they are vacuum-tight (especially if gases are dangerous). 

Pump should be connected to inlet line without any tension. Use flex lines or pipe 

unions in your inlet and exhaust lines so that they can be easily removed for pump 
maintenance. 

The maximum pressure at the inlet may not exceed atmospheric pressure (about 1013 

mbar). Never operate the pump in the presence of over pressures at its intake.  Type of 
materials used for mounting of canalizations should take care of pumped gases. It is the 
same for its tightness. 

If the process gas contains dust or particles, it is absolutely essential to install a dust 

filter in addition to the dust trap. We recommend to install the dust filter horizontally or the 
elbow. This ensures that when removing the filter, no particle falls into the intake port. 

When pumping vapors, we recommend installing condensate traps or condensers on 

the intake side 

By pumping of dangerous gases, inlet line must be tight.   

No particles or liquids may enter in the pump. 

 

Содержание SRV300

Страница 1: ...Instruction Manual VERSION 1 0 Single stage oil sealed rotary vane vacuum pump SRV300...

Страница 2: ...ping of Condensable Gases and Vapors 9 6 7 3 Cyclic process 10 6 8 Stop 11 6 9 Ultimate pressure 11 7 Maintenance 12 7 1 Maintenance schedule 12 7 2 Checking the oil 13 7 2 1 Oil level 13 7 2 2 Oil co...

Страница 3: ...or ordinances Warning Follow the safety instructions below and take appropriate precautions Otherwise it may cause personal injury and equipment damage All power sources connected to the pump must be...

Страница 4: ...ith explosion danger to avoid injury or fire Do not place flammable materials around the motor and the pump body to prevent fire Do not place any obstructions that obstruct ventilation around the moto...

Страница 5: ...n fine filtering occurs in the exhaust filter elements The proportion of oil in the exhaust gas is thus reduced below the visibility threshold over 99 entrapment rate Oil trapped in the exhaust filter...

Страница 6: ...pump Pumps are not designed for pumping of aggressive corrosive flammable or explosive gases These pumps are not designed for working in flammable or explosive environment The pump is not suitable fo...

Страница 7: ...5 Installation diagram 7...

Страница 8: ...e line is too narrow it reduces the pumping speed We recommend applying either LOCTITE or TEFLON tape to the screwed unions so that they are vacuum tight especially if gases are dangerous Pump should...

Страница 9: ...overpressure may occur in the pump Install the exhaust line with a downward slope to prevent condensate from flowing back into the pump If this is not possible we strongly recommend to install a conde...

Страница 10: ...ician in accordance with the applicable safety regulations Wire the motor for the correct supply voltage via connections in the junction box See the wiring diagram on the motor name plate For proper c...

Страница 11: ...the intake to check vacuum Exposure of a part of the body to the vacuum result in a rush of blood in the exposed part 6 7 Operation The pump gases and vapors provided that the gas ballast valve is ins...

Страница 12: ...le vapors until it has warmed to operating temperature pumping process gas with a cold pump results in vapors condensing in the oil For processes with a high proportion of condensable vapors slowly op...

Страница 13: ...ust be cooled down to the ambient temperature and must only be switched on again manually after having eliminated the cause In order to prevent the pump from running up unexpectedly after a main power...

Страница 14: ...nt conditions and or aggressive media may necessitate more frequent maintenance Maintenance job Frequency Section Check the oil level daily 7 2 1 Check the oil condition Depends on process 7 2 2 1st o...

Страница 15: ...operating temperature Depending on the process involved dangerous substances may escape from the pump and oil Take the appropriate precautions 7 3 Oil Change Always change the oil when the pump is sw...

Страница 16: ...that the filter element spring is behind the W shaped filter element bracket 6 Slowly insert the two positioning pins on the exhaust port cover into the corresponding holes in the W shaped filter ele...

Страница 17: ...mm away from the bottom side of the intake port Depending on the process involved dangerous substances may escape from the pump and oil Take the appropriate precautions 7 7 Radiator cleaning Please ke...

Страница 18: ...he pump The gas ballast must be closed and if the pump is to be shelved for a longer period of time it should be sealed in a plastic bag together with a desiccant If the pump has been shelved for more...

Страница 19: ...red without our company s consent 5 Failures and defects caused by abnormal working environment strong electromagnetic radioactive environment high temperature high humidity flammable gas environment...

Страница 20: ...ernal leak 1 Piping fittings loose Repair the pump Retighten Anti suck back valve is malfunctioning Repair the valve Exhaust valve is malfunctioning Repair the valve Float valve does not close Repair...

Страница 21: ...n intake line or in vacuum vessel Oil comes from the vacuum system Check the vacuum system Anti suck back valve is obstructed Clean or repair the valve Sealing surfaces or anti suck back valve are dam...

Страница 22: ...l elapsed time about 24hr continuous operation Intermittent operation Whether exposure to pollutants Yes No Name and Chemical formula pollutants Nature and character of pollutants Cleaning solvents an...

Страница 23: ...uipment Co Ltd Add No 55 Juchao Road Jiangkou subdistrict Fenghua District Ningbo City Zhejiang Province China Post code 315500 Tel 0574 88569588 59516555 Fax 0574 88569596 Email bsvac cnbaosi com Htt...