Eshop:
Email:
w
w
w
.e
M
A
R
eg
.c
z
20
RANGE
Accuracy
1) 2)
50Hz ~100Hz
600.0A
1.5% + 5d
100Hz ~400Hz
600.0A
2.0% + 5d
1)
Induced error from adjacent currentcarrying conductor:
<0.1A/A for Models 037, 035
<0.6A/A for Models 033, 031
2)
For Models 033, 031, specified accuracy is for measurements made at the jaw
center. When the conductor is not positioned at the jaw center,add 2%to specified
accuracy for position errors
RANGE
Accuracy
1) 2)
600.0A
2.0% + 5d
1)
Induced error from adjacent currentcarrying conductor: <0.1A/A
2)
Specified with DCZero mode applied to offset the nonzero residual readings, if any
Function
Sensitivity
1)
(Sine RMS)
Range
600V
50V
5.00Hz ~ 999.9Hz
Accuracy: 1%+5d
1)
DCbias, if any, not more than 50% of Sine RMS