I-26/26R Incubator Shakers M1324-0050
User’s Guide
1.1
Front View
This drawing is for reference, to acquaint you with the user interface controls:
Figure 1: Front View
1.2 Specifications
These specifications assume a maximum load of 34 pounds (15.5 kg), including
platforms, clamps, glassware and contents.
I-26/26R Incubator Shaker
SHAKING
Speed
Control Accuracy
Indication
Stroke/Orbit
Single unit: 25-400 rpm
2 or 3 units stacked: 25-250 rpm
±
1 rpm
3 Digit LED, in 1 rpm increments
1-inch (2.5 cm)
TEMPERATURE
I-26 Range
I-26R Range
Control Accuracy
Indication
Heaters
5°C above ambient temperature to 60°C
15°C below ambient (minimum 4°C) to 60°C
±
0.1°C from 30-40 °C,
±
0.5°C over the remaining range.
3 Digit LED, in 0.1°C increments
Long-life, low-watt density resistance-type heaters with high
temperature thermostats
ALARMS
Visible and audible warning indication when speed deviates more
than 5 rpm or temperature more than 1°C from setpoints, and when
timer has expired. Audible alarm can be muted.