Operation
34
Z33119_1_003
7.5
Remote Operation Modes on BVT
Rotary switch
TopSpin EDTE
Operation
Notes
Remote
Off
No cooling power
This setting is used when VT gas at room temperature has to be
fed to the NMR probe or the VT gas tubing of the BCU I has to be
dried and purged. In this setting the compressor is not running.
Purge with dry clean air or nitrogen gas at room temperature with
a dew point temperature ≤ -50°C@1bar
Remote
On
Strong cooling power
In this maximum mode, the VT gas is cooled to the lowest
specified temperature.