95 (167)
B-ACS RS232C Serial Link
11.5
In its standard configuration the B-ACS sample changer is able to handle 60 or
120 samples, and is controlled via a RS232C serial link. The sample changer soft-
ware defines the transmission mode of the RS232 channel as follows:
8 data bits (data bit 8 = „1“)
2 stop bits
no parity
System Parameters
11.5.1
The following system parameters have to be adjusted or taken in consideration:
1. Check the DIL switch setting on the BMP SIOA board. Switch S1 serves for
setting the I/O address of the SIO board. It should be set at E. The transmis-
sion baud rate can be set with switch S2 according to the following table:
The standard value is 9600 baud, i.e. position 9. The location of the switches are
shown in the implementation diagram,
(Factory settings: S1 = „E“, S2
= „9“).
Table 11.6.
Transmission Baud Rate for RS232
Baud Rate
Switch
Position
50
0
75
1
110
2
150
3
300
4
600
5
1200
6
2400
7
4800
8
9600
9
19200
A to F
Содержание B-ACS 120
Страница 1: ...BRUKER Version Bruker Automated Control Systems B ACS 60 120 Installation Manual 001...
Страница 7: ...Installation Manual Version 001 BRUKER 7 167 1 Declaration of Conformity 1...
Страница 10: ...10 167 BRUKER Installation Manual Version 001 Introduction...
Страница 40: ...40 167 BRUKER Installation Manual Version 001 Operating Instructions...
Страница 56: ...56 167 BRUKER Installation Manual Version 001 Special Tools...
Страница 60: ...60 167 BRUKER Installation Manual Version 001 Technical Data...
Страница 76: ...76 167 BRUKER Installation Manual Version 001 Circuit Diagrams Figure 11 13 Interface Board Circuit Diagram...
Страница 78: ...78 167 BRUKER Installation Manual Version 001 Circuit Diagrams Figure 11 15 Display Board Circuit Diagram...
Страница 82: ...82 167 BRUKER Installation Manual Version 001 Circuit Diagrams Figure 11 18 B ACS Opto Receiver Board...
Страница 124: ...124 167 BRUKER Installation Manual Version 001 Barcode Reader Option Figure 12 13 Barcode Detection Board Digital 2...
Страница 127: ...Schematic Diagrams Installation Manual Version 001 BRUKER 127 167 Figure 12 17 Barcode Adapter Board B ACS 120...
Страница 150: ...150 167 BRUKER Installation Manual Version 001...
Страница 160: ...160 167 BRUKER Installation Manual Version 001 Figures...
Страница 167: ...Installation Manual Version 001 BRUKER 167 167 Lastpage...