Basic Setup Procedures
User Manual Version 002
BRUKER BIOSPIN
75 (327)
For optimum shims (rarely required) set the shims x, y, z
2
, xy and x
2
-y
2
, as well as
xz and yz. You may need to increase the acquisition time
aq
to see the effect of
increasing resolution. Save the shims using the command
wsh
followed by a suit-
able name.
Before the shims are saved, it is recommended to reset the field value in the
bsmsdisp
menu to be exactly on resonance with your shift reference sample of
choice (protons of adamantane or water in D
2
O or silicon rubber).
This allows the command
probefield
(TopSpin version 3 and up) to set the field
according to probe shims and magnet drift (see below, and
Note: For long acquisition times (
aq
> 0.05 s) the decoupling power level
pl12
must be set to +3 dB and d1 must be increased to 6s. To allow longer acquisition
times than 50 msec, the
ZGoption –Dlacq
must be set in
ased
, if the pulse pro-
gram contains the include file aq_prot.incl. Make sure the
–Dlacq
option is not left
set for the following steps.
Calibrating Chemical Shifts on Adamantane
4.5
In TOPSPIN (as well as in the XWIN-NMR 3.5 release) the frequency list for NMR
nuclei follows the IUPAC recommendations (see: R.K. Harris, E.D. Becker, S.M.
Cabral de Menezes, R. Goodfellow and P. Granger,
NMR Nomenclature. Nuclear
Spin Properties and conventions for Chemical Shifts
, Pure Appl. Chem. Vol. 73,
1795-1818 (2001) for reference).
Set the
13
C low field signal of adamantane to 38.48 ppm. This will set the param-
eter
SR
which is used to calculate the chemical shift axis and the peak positions in
the spectrum.
Note: All data sets generated from this data set will have the peak positions cor-
rectly calibrated, if the magnetic field B
0
is not changed. However, you must make
sure that the magnetic field is always the same. It may change, if the magnet has
a slight drift, or if different shim settings are loaded. Therefore the same shim file
should be loaded and the field be set to the same value using the
BSMSDISP
command. If the magnet drift is noticeable, the calibration should be redone in
suitable intervals and the
field
value recorded in the lab notebook.
One can also use a spinner filled with H
2
O to set the field position more precisely.
Do not spin the sample and make sure the cap is well fitted. Set
o1p
to 4.85 ppm,
set for proton observe (as described above for adamantane), and use
gs
for con-
tinuous pulsing and FID display. Change the field value in
bsmsdisp
until the FID
is exactly on resonance. Then all spectra taken should be correctly referenced
with
sr
= 0. For more information on correct field setting and shift calibration, see
"Field Setting and Shift Calibration" on page 87
For all these experiments the field sweep must be off! When the BSMS unit is
turned off and on again, the sweep will always be on. Running spectra with the
sweep on will superimpose spectra at different fields! One can set the sweep am-
plitude to 0 in order to avoid such an accidental error condition.
Содержание Solid State NMR
Страница 1: ...Version Bruker BioSpin think forward NMR Spectroscopy Solid State NMR 002 AVANCE Solids User Manual...
Страница 8: ...8 BRUKER BIOSPIN User Manual Version 002 Contents...
Страница 14: ...14 327 BRUKER BIOSPIN User Manual Version 002 Test Samples...
Страница 104: ...104 327 BRUKER BIOSPIN User Manual Version 002 Practical CP MAS Spectroscopy on Spin 1 2 Nuclei...
Страница 118: ...118 327 BRUKER BIOSPIN User Manual Version 002 Basic CP MAS Experiments...
Страница 122: ...122 327 BRUKER BIOSPIN User Manual Version 002 FSLG HETCOR Figure 8 3 The ased Display...
Страница 154: ...154 327 BRUKER BIOSPIN User Manual Version 002 Proton Driven Spin Diffusion PDSD...
Страница 172: ...172 327 BRUKER BIOSPIN User Manual Version 002 SUPER Figure 13 3 The Acquisition Parameter Window eda...
Страница 178: ...178 327 BRUKER BIOSPIN User Manual Version 002 SUPER...
Страница 192: ...192 327 BRUKER BIOSPIN User Manual Version 002 Symmetry Based Recoupling...
Страница 212: ...212 327 BRUKER BIOSPIN User Manual Version 002 Relaxation Measurements...
Страница 270: ...270 327 BRUKER BIOSPIN User Manual Version 002 Double CP...
Страница 284: ...284 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 1D Figure 22 9 WPMLG CRAMPS After Optimization Digital Acquisition...
Страница 290: ...290 327 BRUKER BIOSPIN User Manual Version 002 Modified W PMLG...
Страница 302: ...302 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 2D...
Страница 304: ...304 327 BRUKER BIOSPIN User Manual Version 002...
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 318: ...318 327 BRUKER BIOSPIN User Manual Version 002 Tables...
Страница 326: ...326 327 BRUKER BIOSPIN User Manual Version 002 Index...
Страница 327: ...User Manual Version 002 BRUKER BIOSPIN 327 327 End of Document...