![Bruker BioSpin Solid State NMR Скачать руководство пользователя страница 258](http://html1.mh-extra.com/html/bruker-biospin/solid-state-nmr/solid-state-nmr_user-manual_2821040258.webp)
258 (327)
BRUKER BIOSPIN
User Manual Version 002
Double-CP
6. Optimize decoupling and CP condition, run a reference
13
C CP/MAS spectrum
of the labelled glycine sample, using 16 scans. This reference spectrum will
serve to measure the efficiency of the DCP magnetization transfer.
7. To set up the conditions for the N to C transfer, one must define the RF field at
which the transfer is to take place, and find the appropriate power levels to
achieve these RF fields. In order to minimize losses due to insufficient excita-
tion bandwidths and T
1
ρ
relaxation, the contact should be executed at high
power. However, there are limitations in terms of what the probe can take, and
there are losses due to unwanted HH contact to the proton spin system. On
the other hand, in many samples (bio-samples) the spread of chemical shifts
that one wants to cover is not extremely wide or one even wants to execute the
transfer selectively (“Specific-CP”). An RF field of 35 kHz is a decent compro-
mise. So, using the
cp90
pulse program, and moving the carrier close to the
C
α
-peak, determine a power level
pl11
which corresponds to 35 kHz RF (7.14
μ
sec 90 ° pulse). However, since the sample spins at 11 kHz, the HH condition
will require 46 or 24 kHz on one nucleus and 35 kHz on the other nucleus. If
you decide to account for the spin rate on the
13
C side, calculate the required
power levels for 24 and 46 kHz (= 35 kHz RF field +/- 11 kHz spin rate) RF field
using
xau
calcpowlev
. Now the
13
C channel is set.
15
N Channel Setup
20.2.2
8. Create a new experiment using edc.
9. Setup the proper routing by going into
edasp
. Click the switch F1/F3 button to
get
15
N on channel 1.
Figure 20.3. Routing table for triple resonance setup change for
15
N pulse param-
eter measurement and CPMAS optimization.
10. If available, set
pl1
and
sp0
for a proton/
15
N HH condition in triple mode. On a
labelled sample, even the previous settings for
13
C should give a signal which
allows optimizing the HH condition.
11. When the HH condition is optimized, find the power level to achieve a 35 kHz
RF field (7.14
μ
sec 90 ° pulse, carrier close to the
15
N-resonance). It is essen-
tial to optimize the first proton to nitrogen HH contact. This is not as trivial as
Содержание Solid State NMR
Страница 1: ...Version Bruker BioSpin think forward NMR Spectroscopy Solid State NMR 002 AVANCE Solids User Manual...
Страница 8: ...8 BRUKER BIOSPIN User Manual Version 002 Contents...
Страница 14: ...14 327 BRUKER BIOSPIN User Manual Version 002 Test Samples...
Страница 104: ...104 327 BRUKER BIOSPIN User Manual Version 002 Practical CP MAS Spectroscopy on Spin 1 2 Nuclei...
Страница 118: ...118 327 BRUKER BIOSPIN User Manual Version 002 Basic CP MAS Experiments...
Страница 122: ...122 327 BRUKER BIOSPIN User Manual Version 002 FSLG HETCOR Figure 8 3 The ased Display...
Страница 154: ...154 327 BRUKER BIOSPIN User Manual Version 002 Proton Driven Spin Diffusion PDSD...
Страница 172: ...172 327 BRUKER BIOSPIN User Manual Version 002 SUPER Figure 13 3 The Acquisition Parameter Window eda...
Страница 178: ...178 327 BRUKER BIOSPIN User Manual Version 002 SUPER...
Страница 192: ...192 327 BRUKER BIOSPIN User Manual Version 002 Symmetry Based Recoupling...
Страница 212: ...212 327 BRUKER BIOSPIN User Manual Version 002 Relaxation Measurements...
Страница 270: ...270 327 BRUKER BIOSPIN User Manual Version 002 Double CP...
Страница 284: ...284 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 1D Figure 22 9 WPMLG CRAMPS After Optimization Digital Acquisition...
Страница 290: ...290 327 BRUKER BIOSPIN User Manual Version 002 Modified W PMLG...
Страница 302: ...302 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 2D...
Страница 304: ...304 327 BRUKER BIOSPIN User Manual Version 002...
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 318: ...318 327 BRUKER BIOSPIN User Manual Version 002 Tables...
Страница 326: ...326 327 BRUKER BIOSPIN User Manual Version 002 Index...
Страница 327: ...User Manual Version 002 BRUKER BIOSPIN 327 327 End of Document...