![Bruker BioSpin Solid State NMR Скачать руководство пользователя страница 246](http://html1.mh-extra.com/html/bruker-biospin/solid-state-nmr/solid-state-nmr_user-manual_2821040246.webp)
246 (327)
BRUKER BIOSPIN
User Manual Version 002
STMAS
which are indicated by the (red) filled circles. The open rectangles symbolize the
second pulse (one pulse at the end of each individual t
1
increment). They must al-
ways occur precisely on top of the rotational echo. For each increment the t
2
data
acquisition, which is not shown here, starts after the second pulse.
In the following discussion we will ignore this part completely. It showed up in the
original experiments but can be completely suppressed by a double quantum fil-
ter. The contribution of the ST “rides” on top of the CT signal like spikelets. Since
MAS efficiently averages the 1st order quadrupole interaction of the ST, the corre-
sponding MHz broad signal is now narrowed into a huge number of spinning side
bands. These coherency originating from the ST dephase rapidly and refocus into
rotational echoes with each rotor cycle. A pulse precisely on top of such a rota-
tional echo can transfer the SQ coherency from the ST to SQ coherency of the
CT, the signal from which can then be acquired under standard MAS conditions.
The evolution in the indirect dimension is achieved in such a way that the delay
between the two pulses, which is the evolution period t
1
, is incremented by integer
multiples of the rotor period.
Two extremely important points must be considered for the experimental realiza-
tion of the STMAS experiment. Firstly, the spinning frequency must be kept abso-
lutely constant. The duration of the rotational echoes in the STMAS experiment is
determined by the width of the satellite transition, giving a length of e.g. 1 µs for a
satellite transition of 1 MHz width. If the rotor period varies from that specified in
the parameters, the calculated delay in the pulse program is incorrect and the
pulse misses the echo top, so less or no signal intensity is obtained.
summarizes the time deviation that occurs when the spinning frequency fluctuates
by
±
1 Hz and
±
10 Hz at various desired spinning frequencies. One can see that
when the t
1
increment accumulates to as much as 100 rotor periods it is possible
to miss an echo completely. For example, if the duration of the rotational echo is 1
µs it will be missed when the deviation is larger, which is the case for a 1 Hz devi-
ation at 10 kHz, but requires a fluctuation of 10 Hz at 30 kHz spinning.
Typically the spinning frequency must be stable within
≤
1 Hz throughout the entire
2D data acquisition. Secondly, the accuracy of the magic angle setting is extreme-
ly important. The sidebands resulting from the first order broadening are narrowed
from the full first order line width by a factor of
(3cos
2
θ
-1)
, hence for a deviation of
d
θ
from the magic angle the broadening is
3cos
θ
sin
θ
d
θ
, which close to the magic
angle is
√
2d
θ
. The magnitude of the interaction that must be narrowed in the pres-
Table 19.1.
Time deviation of the rotor period for spinning frequency variations of
±
1 and
±
10 Hz for
various spinning frequencies.
Fluctuation of … Hz
@ desired spinning
frequency
Deviation from precise rotor
period after 1 rotor period
Deviation from precise rotor period
after 100 rotor periods
10 Hz @ 30 kHz
11 ns
1.1 µs
1 Hz @ 30 kHz
1.1 ns
110 ns
10 Hz @ 20 kHz
25 ns
2.5 µs
1 Hz @ 20 kHz
2.5 ns
250 ns
10 Hz @ 10 kHz
100 ns
10 µs
1 Hz @ 10 kHz
10 ns
1 µs
Содержание Solid State NMR
Страница 1: ...Version Bruker BioSpin think forward NMR Spectroscopy Solid State NMR 002 AVANCE Solids User Manual...
Страница 8: ...8 BRUKER BIOSPIN User Manual Version 002 Contents...
Страница 14: ...14 327 BRUKER BIOSPIN User Manual Version 002 Test Samples...
Страница 104: ...104 327 BRUKER BIOSPIN User Manual Version 002 Practical CP MAS Spectroscopy on Spin 1 2 Nuclei...
Страница 118: ...118 327 BRUKER BIOSPIN User Manual Version 002 Basic CP MAS Experiments...
Страница 122: ...122 327 BRUKER BIOSPIN User Manual Version 002 FSLG HETCOR Figure 8 3 The ased Display...
Страница 154: ...154 327 BRUKER BIOSPIN User Manual Version 002 Proton Driven Spin Diffusion PDSD...
Страница 172: ...172 327 BRUKER BIOSPIN User Manual Version 002 SUPER Figure 13 3 The Acquisition Parameter Window eda...
Страница 178: ...178 327 BRUKER BIOSPIN User Manual Version 002 SUPER...
Страница 192: ...192 327 BRUKER BIOSPIN User Manual Version 002 Symmetry Based Recoupling...
Страница 212: ...212 327 BRUKER BIOSPIN User Manual Version 002 Relaxation Measurements...
Страница 270: ...270 327 BRUKER BIOSPIN User Manual Version 002 Double CP...
Страница 284: ...284 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 1D Figure 22 9 WPMLG CRAMPS After Optimization Digital Acquisition...
Страница 290: ...290 327 BRUKER BIOSPIN User Manual Version 002 Modified W PMLG...
Страница 302: ...302 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 2D...
Страница 304: ...304 327 BRUKER BIOSPIN User Manual Version 002...
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 318: ...318 327 BRUKER BIOSPIN User Manual Version 002 Tables...
Страница 326: ...326 327 BRUKER BIOSPIN User Manual Version 002 Index...
Страница 327: ...User Manual Version 002 BRUKER BIOSPIN 327 327 End of Document...