Introduction
E1400011_1_001
25
Figure 5.2: The minispec Magnet Unit (left) and Electronic Control Unit (right).
A number of optional accessories can be added and adapted for use: Gradient Unit (for
pulsed field gradient experiments), electronic balances, auto sampler, temperature control
baths and variable temperature control units, etc.
5.3.1
Unpacking the System
When unpacking the instrument ensure that all components listed in the packing list are part
of the delivery.
CAUTION
Danger of property damage due to improper handling.
The shipping container must remain in an upright position at all times. It has two
compartments separated by a wooden plate: an upper part, which contains the Electronic
Control Unit, and a lower part containing the Magnet Unit. The magnet unit is secured to the
shipping container from below with four bolts.
u
DO NOT tilt the container when removing these bolts.
CAUTION
Accident or material damage hazard from falling or tilting instrument
Instrument may fall down or tilt during installation, retrofitting, dismantling, or while moving
on a cart.
u
Use the carrying handles during lifting.
u
When a table or cart is used, it should meet the requirements in the site planning guide.
u
The instrument should be secured to the cart or table.
Содержание minispec mq10
Страница 1: ...minispec minispec mq Series User Manual Version 001 Innovation with Integrity AIC ...
Страница 12: ...Safety 12 E1400011_1_001 ...
Страница 16: ...Transport Packaging and Storage 16 E1400011_1_001 ...
Страница 66: ...Maintenance 66 E1400011_1_001 ...
Страница 68: ...Dismantling and Disposal 68 E1400011_1_001 ...
Страница 70: ...Contact 70 E1400011_1_001 ...
Страница 72: ...List of Figures 72 E1400011_1_001 ...
Страница 75: ...E1400011_1_001 75 ...
Страница 76: ...Bruker Corporation info bruker com www bruker com Order No E1400011 ...