CHAPTER 6
Specifications
29
Compliance with Standards
,
CE-mark indicates compliance with: EMC Directive and Low Voltage Directive.
C-Tick mark indicates compliance with the EMC requirements of Australia and New
Zealand
Safety
EN 61010
−
1 and IEC 61010
−
1: Safety requirements for electrical equipment for
measurement, control and laboratory use.
UL 3111
−
1: Standard for Safety
−
Electrical measuring and test equipment
EMC Emission
EN 61326
−
1: Electrical equipment for measurement, control and laboratory use.
EMC requirements.
Part 1: General requirements.
EN 50081– 1: Generic emission standard. Part 1: Residential, commercial and light
industry.
EN 50081– 2: Generic emission standard. Part 2: Industrial environment.
CISPR 22: Radio disturbance characteristics of information technology equipment.
Class B Limits.
FCC Rules, Part 15: Complies with the limits for a Class B digital device.
EMC Immunity
EN 61326
−
1: Electrical equipment for measurement, control and laboratory use.
EMC requirements.
Part 1: General requirements.
EN 50082– 1: Generic immunity standard. Part 1: Residential, commercial and light
industry.
EN 50082– 2: Generic immunity standard. Part 2: Industrial environment.
ISO 7637– 1, 7637
−
2 and 7637– 3: Road Vehicles — Electrical Disturbance by
Conduction and Coupling.
Note 1: Refer to “Environmental Susceptibility” in specifications.
Note 2: The above is guaranteed using the accessories in this PD sheet only..
Temperature
IEC 60068– 2– 1 & IEC 60068– 2– 2: Environmental Testing. Cold and Dry Heat.
Operating Temperature: – 10 to +55
°
C (14 to 131
°
F)
Storage Temperature: –25 to +70
°
C (–13 to 158
°
F)
IEC 60068– 2– 14: Change of Temperature: –10 to +55
°
C (2 cycles, 1
°
C/min.)
Humidity
IEC 68– 2– 3: Damp Heat: 90% RH (non-condensing at 40
°
C (104
°
F))
Mechanical
Operating (peak values)
MIL– STD – 810C: Vibration: 12.7 mm, 15 m/s
2
, 5 – 500 Hz
Non-operating:
IEC 60068– 2– 6: Vibration: 0.3 mm, 20 m/s
2
, 10– 500 Hz
IEC 60068– 2– 27: Shock: 1000 m/s
2
IEC 60068– 2– 29: Bump: 1000 bumps at 250 m/s
2
Enclosure
IEC 60529 (1989): Protection provided by enclosures: IP20
BE164411.book Page 29 Friday, May 18, 2001 11:51 AM
Содержание 2694A
Страница 3: ...BE 1644 11 May 2001 BE164411 book Page 2 Friday May 18 2001 11 51 AM...
Страница 5: ...BE164411 book Page 4 Friday May 18 2001 11 51 AM...
Страница 7: ...BE164411 book Page 4 Friday May 18 2001 11 51 AM...
Страница 15: ...16 channel Conditioning Amplifier Type 2694 User Manual 8 BE164411 book Page 8 Friday May 18 2001 11 51 AM...
Страница 21: ...16 channel Conditioning Amplifier Type 2694 User Manual 14 BE164411 book Page 14 Friday May 18 2001 11 51 AM...
Страница 31: ...16 channel Conditioning Amplifier Type 2694 User Manual 24 BE164411 book Page 24 Friday May 18 2001 11 51 AM...
Страница 37: ...16 channel Conditioning Amplifier Type 2694 User Manual 30 BE164411 book Page 30 Friday May 18 2001 11 51 AM...
Страница 41: ...16 channel Conditioning Amplifier Type 2694 User Manual 34 BE164411 book Page 34 Friday May 18 2001 11 51 AM...