16-channel Conditioning Amplifier Type 2694 –

User Manual

16

The 

Help 

button 

 in the taskbar is a useful function. To use it, simply right-click

on the button in the task bar, release, and right-click again on the area you would
like explained. This Help option works both on- and off-line.

Example Setup: 

Attach some transducers to a Type 2694 amplifier (assume that the

system is set up to measure DeltaTron transducers). At the input sockets, the left-
hand (red) LED switches off, leaving a single, green LED lit. This shows that a trans-
ducer is attached and ready for measuring. To identify those transducers with TEDS,
and with the Transducer Setup tab clicked, go to the taskbar and click on the

Update IDs

 button 

. The Setup view automatically updates with the serial

number, type number, sensitivity and units of all attached, TEDS-compatible trans-
ducers, while displaying which input sockets have transducers attached. Transduc-
ers without TEDS need to have their parameters entered manually. 

Channels not in use or not required for measuring can be disabled by right-clicking
on the relevant channel in the 

Channel

 column of the Setup view, and selecting

Disable

 from the drop-down list. The 

Send Parameters

 button must then be clicked

to notify the Type 2694 amplifier. Once a channel has been disabled, all the LEDs
above the corresponding input socket on the instrument are switched off.

Now click on the Amplifier Setup tab. Parameters here are similarly adjusted by
right-clicking on individual cells, choosing an option from the scroll-down list, and
clicking the

 Send Parameters

 button on the task bar afterwards. 

Overloads are covered in the 

Overload

 column where a cell’s colour can be blue or

red. A blue cell implies that there has been no overload (since the last reset), while
a red cell implies that there is a current overload. A small, black triangle in the top
left-hand corner of a cell implies that an overload has occurred (called a latched
overload) since the last reset. The 

Reset Overloads

 button 

 on the taskbar resets

all overloads. Channels that have been disabled or are not in use have cells col-
oured blue.

To see when overloads have occurred, go to the Menu bar and click 

View

Message

Log

 to get a window displaying overload time, port number, amplifier number, chan-

nel, error code and description, i.e., a full history of the RS–232 link.

The File Manager view can be used on- or off-line, but the latter allows you to create
different setups for Type 2694 amplifier. In the Menu bar, click 

File

New

. From Link

view, drag a previous setup and drop it into your new data file in the File Manager
view. Now, drag this file onto the Setup view and proceed to edit the setup. Once
edited, right-click on the file and save. You can then drag the file back to the Link
view and drop it on the instrument icon. Clicking on the 

Send Parameters

 button in

the task bar loads the new setup to the instrument. 

Amplifier Setup Menu

The parameters in the 

Amplifier Setup

 menu directly affect the measured output

and must therefore be set up before measuring.

BE164411.book  Page 16  Friday, May 18, 2001  11:51 AM

Содержание 2694A

Страница 1: ...600 Italy 02 57 68061 Japan 03 3779 8671 Republic of Korea 02 3473 0605 Netherlands 31 318 559290 Norway 66771155 Poland 22 858 9392 Portugal 1 4711453 Singapore 65 377 4512 Slovak Republic 421 7 544...

Страница 2: ...BE 1644 11 May 2001 16 channel Conditioning Amplifier Type 2694 User Manual Valid for version 1 0 or higher of the Type 2694 Embedded Software BE164411 book Page 1 Friday May 18 2001 11 51 AM...

Страница 3: ...BE 1644 11 May 2001 BE164411 book Page 2 Friday May 18 2001 11 51 AM...

Страница 4: ...llowing Safety Symbols The apparatus will be marked with this symbol when it is important that you refer to the associated warning statements given in the manual Protective Earth Terminal Hazardous Vo...

Страница 5: ...BE164411 book Page 4 Friday May 18 2001 11 51 AM...

Страница 6: ...and Connections 9 Front Panel 9 EMC Immunity 10 CHAPTER 3 Operation Overview 11 Preliminary 11 CHAPTER 4 The Use of Control Software BZ 5291 15 Introduction 15 Transducer Setup 20 CHAPTER 5 Maintenanc...

Страница 7: ...BE164411 book Page 4 Friday May 18 2001 11 51 AM...

Страница 8: ...lways be updated changed to another configuration if required but for calibration and reliability reasons the unit must first be returned to Br el Kj r for updating Note Type 2694 B cannot be upgraded...

Страница 9: ...d double integration on all channels 000280 1 Floating Single ended Input Input Protection Voltage Mode 6mA Transducer Supply IEEE 1451 4 0 1Hz H p 1Hz H P Disable 10dB 0dB 10dB 20dB 30dB 40dB Optiona...

Страница 10: ...in 0 dB Gain 10 dB Gain 20 dB Gain 30 dB Gain 40 dB IEPE DeltaTron Input Voltage Input IEEE P1451 4 Transducer Support Optional Filters Possible Yes No Yes Yes Filters Installed e g A B C D or single...

Страница 11: ...tioning amplifiers is automatically detected by the software and displayed in the File and Link views see Fig 1 4 You designate the port s used for the range of conditioning amplifiers yourself You ca...

Страница 12: ...r setup see Fig 1 5 you can specify the settings of filters and gain for each channel This includes high pass filters optional filters gain in steps of 10 dB multiplexer channel tachometer and whether...

Страница 13: ...is only for use in the analyzers or data acquisi tion systems accompanying the conditioning amplifier Sensitivities from other non TEDS transducers cannot be entered manually in the Sensitivity field...

Страница 14: ...one COM port on a PC Each unit is automatically addressed from the PC program Setup and Control software BZ 5291 which is included with each Type 2694 unit One other serially controlled instrument fo...

Страница 15: ...16 channel Conditioning Amplifier Type 2694 User Manual 8 BE164411 book Page 8 Friday May 18 2001 11 51 AM...

Страница 16: ...sing the control software DC 10 33 V Coaxial DC socket for connection to an external 10 to 33 V DC source Complies with ISO DP 7637 1 and DP 7637 2 24 V Warning The DC socket cable ground is not conne...

Страница 17: ...t is used in elec trically hostile environments such as powerful electrical fields In such cases the cables for the power supply the RS 232 interface and all floating inputs outputs must be fitted wit...

Страница 18: ...eet the strict ISO 7637 1 Road Vehicles Electrical disturbance by conduction and coupling requirements Power Supply External Power The conditioning amplifier is powered from an external DC supply whic...

Страница 19: ...n racks For automotive use a special portable 19 rack KQ 0158 is available Grounding Ground loops are a common problem in large vibration measurement installations There are different ways of avoiding...

Страница 20: ...ers are connected to the same ground structure In such setups magnetic fields will induce current into the loop causing noise signals In such cases the best solution is to select floating inputs excep...

Страница 21: ...16 channel Conditioning Amplifier Type 2694 User Manual 14 BE164411 book Page 14 Friday May 18 2001 11 51 AM...

Страница 22: ...ponding to individual channels and the columns corresponding to different channel parameters Column width is adjustable by dragging the column edges with the mouse and extra columns can be inserted an...

Страница 23: ...y adjusted by right clicking on individual cells choosing an option from the scroll down list and clicking the Send Parameters button on the task bar afterwards Overloads are covered in the Overload c...

Страница 24: ...e Integration The acceleration signal is integrated once to provide a velocity output signal This is the instantaneous vibration velocity magnitude Double Integration The acceleration signal is integr...

Страница 25: ...alid The overload conditions listed in Table 4 3 are indicated by the red LEDs on the front panel for the respective channels The LEDs indicate instantane ous overload conditions Alternatively the ove...

Страница 26: ...ansducer Current Overload Another type of overload not detected by Condition ing Amplifier Type 2694 is the transducer current overload This occurs when exces sive drive current is required based on t...

Страница 27: ...sensitivity and measurement units in this setup menu When a channel is disabled its LEDs are not lit Note Changing the sensitivity has no effect on the amplification it is only informa tion for the us...

Страница 28: ...al Fig 4 2 Connections of the BNT socket for DeltaTron input When used with BNC cables the connector will function as a normal BNC connector Voltage input mode is used for general voltage input signal...

Страница 29: ...ccelerometer 1 1 1 4 Accelerometer transfer function 65 1 1 5 Charge amplifier accelerometer 321 2 2 6 Charge amplifier transfer function accelerometer 577 2 2 7 Microphone integrated preamplifier 769...

Страница 30: ...hed Software Version Shows the version number for the boot and application software that constitutes the current Type 2694 program These numbers are only changed if the Type 2694 ampli fier is updated...

Страница 31: ...16 channel Conditioning Amplifier Type 2694 User Manual 24 BE164411 book Page 24 Friday May 18 2001 11 51 AM...

Страница 32: ...emicals avoid using strong detergents Be careful not to allow water droplets to enter the ventilation holes Service and Repair Hardware The Type 2694 Range of Conditioning Amplifiers are designed and...

Страница 33: ...r it not being possible to turn it on there are a number of possible causes and solutions l Heavily loaded outputs remove loads l DC supply voltage too low 10 V at the Type 2694 power input check supp...

Страница 34: ...nnel to channel Phase Match calculated values 1 flow 2 degrees 10 flow 0 25 degrees 100 flow 0 025 degrees Flexible Filter Configuration Built in filters High pass filters versions A C and D There are...

Страница 35: ...TAL CONTROL INTERFACE Serial Interface RS 232 Computer Control All functions are controlled via the RS 232 interface You can daisy chain up to 16 units on each COM port Support of Transducers with TED...

Страница 36: ...t for measurement control and laboratory use EMC requirements Part 1 General requirements EN 50082 1 Generic immunity standard Part 1 Residential commercial and light industry EN 50082 2 Generic immun...

Страница 37: ...16 channel Conditioning Amplifier Type 2694 User Manual 30 BE164411 book Page 30 Friday May 18 2001 11 51 AM...

Страница 38: ...n 12 Double integration Filters 17 Drive current 19 E Electromagnetic compatibility 11 EMC immunity 10 29 Environmental Electromagnetic compatibility 11 Noise 17 Susceptibility 28 Temperature 11 29 Ex...

Страница 39: ...ad Indications 18 Internal signal 16 20 Voltage 19 P Peak meter menu 18 Phase errors 17 Potential difference 12 Power Consumption 11 External 11 28 Supply 11 R Rack mounting 12 Repair 25 Revision stat...

Страница 40: ...INDEX 33 V Version number 23 Voltage Bias 19 DC supply 11 Mains range 12 BE164411 book Page 33 Friday May 18 2001 11 51 AM...

Страница 41: ...16 channel Conditioning Amplifier Type 2694 User Manual 34 BE164411 book Page 34 Friday May 18 2001 11 51 AM...

Страница 42: ...600 Italy 02 57 68061 Japan 03 3779 8671 Republic of Korea 02 3473 0605 Netherlands 31 318 559290 Norway 66771155 Poland 22 858 9392 Portugal 1 4711453 Singapore 65 377 4512 Slovak Republic 421 7 544...

Отзывы: